User Manual
101 
16 
17 
18 
+5V 
Internal +5V: It is not recommended to use the 
internal +5 V. If this voltage output is used, the 
current should be lower than 0.3 A, and that the 
signal line is far from a disturbance source.   
19 
/PHI 
Indicates that the test result is greater than the high 
limit set on BIN1 to BIN9.  
20 
/PLO 
The main parameter is lower than the lower limit in 
BIN1 to BIN9.  
21 
/SREJ 
The test result is outside the valid range. 
22-24 
NC 
Not connection. 
25 
/KEY LOCK 
When this line is asserted, the keys in the front panel 
are locked.  
27 
28 
/EXT.DCV1 
The external DC voltage 1: this is a pull-up DC Voltage 
pin for optoelectronic coupling signal  
(/BIN-/BIN9, /AUX, /OUT, /PHI, /PLO,/SREJ). 
29 
/ALARM 
When circuit is interrupted, this signal is asserted.  
30 
/INDEX 
When the analog test is finished and the UNKNOWN 
terminal can be connected to another DUT, /INDEX is 
asserted. However, the comparison signal is asserted 
until /EOM is asserted.  
31 
/EOM 
End Of Measurement: when the test data and the 
compared results are asserted, this signal is also 
asserted.  
32 
33 
COM2 
The reference ground for external power EXTV2.  
34-36 
COM1 
The reference ground for external power EXTV1. 
Table 21 - Comparison Output Pinout 
PHI(OUT) 
SREJ(AUX) 
BIN 1 
SREJ(AUX) 
BIN 2 
BIN 3 
BIN 4 
BIN 5 
BIN 6 
BIN 7 
BIN 8 
BIN 9 
PLO(OUT) 
The secondary parameter 
The main 
parameter 