User Manual
61
1. From the SYSTEM menu, select the COMM SETUP function key.
2. From the COMM SETUP menu, press the down arrow key multiple
times to highlight the IP Mode parameter. You have the option
between AUTO and MANU. Selecting AUTO will allow the unit to
request an IP Address, Subnet Mask, and Gateway automatically in
order to communicate properly. No other configuration is necessary
once AUTO is selected.
NOTE: The unit must already be connected via Ethernet connection
and the connected network/router must be compatible to assign
the correct parameters to the instrument.
Selecting MANU will allow the user to manually enter the IP Address,
Subnet Mask, and Gateway of the instrument.
3. If MANU is selected, press the down arrow key to highlight the IP
Address parameter and enter using the numeric keypad (i.e. IP
Address: 192.168.100.55).
4. Repeat step 3 for the Subnet Mask and Gateway parameters.
5. Press ESC to exit editing parameters.
Web Server
There is an embedded web server GUI that can access the LCR meter via LAN
interface using a web browser. The GUI provides a simple way of setting
measurement parameters, sending SCPI commands, making measurements,
as well as viewing system information using a web browser from a computer
connected to the same local area network as the instrument. To access this,
perform the following steps:
1. Open up a web browser on the computer
2. Check the IP address of the LCR meter by pressing the SYSTEM
function key and the SYSTEM INFO key.