User Manual
40
4. Use the arrow keys to edit the desired measurement parameters
such as level, start frequency, stop frequency, high and low values of
the Y-Axis (if AUTOSCALE is ON, the meter will automatically scale the
Y-axis), and the step value. Step can be set to 1, 2, 5, and 10 points
per sweep step.
5. Turn AUTOFREQ ON if you would like the meter to automatically
select the start and stop frequency.
6. Turn INTERP ON if you would like to have interpolated data included
in the sweep result (only valid if step is 2, 5, or 10). Connect the test
leads to the component under test or insert component into the test
fixture.
7. Press the START softkey to execute the sweep function with the
current parameters. The display will do an initial scan and analyze
the component connected to the input to determine the frequency
sweep range (if AUTOFREQ is ON). The below screen will display
during the scan.
8. If needed, press the STOP softkey to terminate the sweep function
immediately.
9. While sweeping, the instrument may occasionally display the below
screen if a measurement requires more than 1000 ms to complete.
This is normal.