User Manual
39
AUTOSCALE
If enabled (ON), the unit will automatically
select the HI and LO values (Y-axis) to properly
display the frequency sweep on the display
graph. If disabled (OFF), the manually set user
values will be used.
ON, OFF
AUTOFREQ
If enabled (ON), the unit will automatically
select the start and stop frequency based on
an initial analysis of the component connected
to the input and the initial start and stop
frequency that is set. The start and stop
frequency are then adjusted automatically to
ensure sufficient interval/spacing between
each frequency points of the sweep.
ON, OFF
INTERP
If enabled (ON), the unit will show interpolated
data in table view and return interpolated data
in remote mode.
ON, OFF
The following table specifies the number of points generated with
interpolation enabled and disabled.
Table 3.9 - Step Parameter Points
Step
Parameter
Number of Points
(INTERP ON)
Number of Points
(INTERP OFF)
1
300
300
2
300
150
5
300
60
10
300
30
To run the sweep function, perform the following steps:
1. Press the SWEEP function key to enter the sweep function.
2. Press the DISP function key to select linear or logarithmic scale.
3. Press the SWAP softkey if the secondary value is desired to be swept.