User Manual

27
will increment by one with beep function. The beep can be turned off or used
when the component passes or fails specification. The beep tone for each bin
is different so that the user can distinguish bins by listening. The bin detection
sequence is from low to high. This means that the bin 1 comparator will
execute first and if the measured value is not within bin 1’s limits, it will
execute the bin 2 comparator and so on. For instance, if both bins are set to
the same limits, the larger numbered bin will not be compared because the
smaller numbered bin will always count first.
To measure the value and record the count into the predefined bins, users
can connect the component and press the TRIG softkey. If the flash drive is
inserted and the Bin Log option is ON in the SYSTEM SETUP menu (see System
Setup section), each measured value will be recorded onto the flash drive.
The bin function has two modes: absolute mode and tolerance mode.
Absolute Mode
The absolute mode allows users to use absolute values for the high and low
limits of each bin. The following example shows a typical setup of absolute
mode.
Figure 3.15 – Bin Mode Table View (Absolute Mode)
The following graph is a visual representation of the Bin absolute mode setup
demonstrated in Figure 3.16.