User Manual
23 
Parameter Combinations 
The combinations of primary and secondary parameters, including the series 
and parallel combinations are listed below: 
Table 3.4 – Parameter Combinations 
Primary  
Parameter 
Series Mode 
Parallel Mode 
C 
Cs-Q 
Cs-D 
Cs-R 
Cp-Q 
Cp-D 
Cp-R 
Cp-G 
L 
Ls-Q 
Ls-D 
Ls-R 
Lp-Q 
Lp-D 
Lp-R 
Lp-G 
Z 
Z-θ 
- 
Y 
- 
Y-θ 
R 
R-X 
- 
G 
- 
G-B 
Measurement Options 
The following table lists the measurement parameter options that are 
configurable for a basic measurement. 
Table 3.5 – Measurement Parameter Options 
Parameter 
Description 
Input 
FUNC 
Specify the measurement type 
All combinations  
FREQ 
Specify the test signal frequency  
20 Hz to 300 kHz 
LEVEL 
Specify the test signal level 
0.5 VRMS, 1 