User Manual

23
Parameter Combinations
The combinations of primary and secondary parameters, including the series
and parallel combinations are listed below:
Table 3.4 – Parameter Combinations
Primary
Parameter
Series Mode
Parallel Mode
C
Cs-Q
Cs-D
Cs-R
Cp-Q
Cp-D
Cp-R
Cp-G
L
Ls-Q
Ls-D
Ls-R
Lp-Q
Lp-D
Lp-R
Lp-G
Z
Z-θ
-
Y
-
Y-θ
R
R-X
-
G
-
G-B
Measurement Options
The following table lists the measurement parameter options that are
configurable for a basic measurement.
Table 3.5 – Measurement Parameter Options
Parameter
Description
Input
FUNC
Specify the measurement type
All combinations
FREQ
Specify the test signal frequency
20 Hz to 300 kHz
LEVEL
Specify the test signal level
0.5 VRMS, 1