Model 830C, 890C Dual Display Capacitance Meter INSTRUCTION MANUAL
Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be observed during all phases of operation, service, and repair of this instrument. DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE Do not operate the instrument in the presence of flammable gases or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard. KEEP AWAY FROM LIVE CIRCUITS Instrument covers must not be removed by operating personnel.
service and repair to ensure that safety features are maintained. WARNINGS AND CAUTIONS WARNING and CAUTION statements, such as the following examples, denote a hazard and appear throughout this manual. Follow all instructions contained in these statements. A WARNING statement calls attention to an operating procedure, practice, or condition, which, if not followed correctly, could result in injury or death to personnel.
SAFETY GUIDELINES To ensure that you use this device safely, follow the safety guidelines listed below: This meter is for indoor use, altitude up to 2,000 m. The warnings and precautions should be read and well understood before the instrument is used. When measuring in-circuit components, first de-energize the circuits before connecting to the test leads. Never attempt to measure voltage with this instrument. Do not use the device if it appears damaged in form.
Category II (CAT. II) 50 V, Pollution Degree 2 environment. Use the meter only as specified in this manual. Otherwise, the protection provided by the meter may be impaired. The power for the meter is supplied with a single standard 9V battery. It can also operate from AC line using a 12V AC to DC adaptor. If a power adaptor is selected, please be sure to use one with correct AC line input voltage and fulfills the safety requirements of a relevant IEC standard.
Safety Symbols This symbol is a warning and indicates that the user should refer to the operating instructions located in the manual.
This product is subject to Directive 2002/96/EC of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE) , and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste. Please utilize your local WEEE collection facilities in the disposition of this product and otherwise observe all applicable requirements.
Environmental Conditions 0 °C to 40 °C Operating Environment Storage Humidity Storage Environment Pollution degree 0 – 80% R.H.
TABLE OF CONTENTS SAFETY GUIDELINES ............................................... 3 INTRODUCTION....................................................... 11 PACKAGE CONTENTS ........................................... 13 FRONT PANEL OVERVIEW .................................... 14 Front Panel Display Descriptions .................................. 15 Front Panel Buttons ...................................................... 16 LCD DISPLAY OVERVIEW ......................................
Baud Rate (bAUd) ........................................................ 33 Parity (PArt) .................................................................. 35 Data Bits (Data) ............................................................ 36 Echo (ECHO) ................................................................ 37 Print Only (Prnt) ............................................................ 38 Key Beeper (bEEP) ...................................................... 39 Lock Buttons (LbUt) ............
Remote Mode ............................................................... 76 Command Protocols ..................................................... 77 MEASUREMENT TIPS ............................................. 97 SUPPLEMENTAL INFORMATION .......................... 99 Guard Terminal ............................................................. 99 SPECIFICATIONS .................................................. 100 General specification ..................................................
INTRODUCTION B&K Precision’s 830C and 890C 11,000-count handheld capacitance meters measure capacitance and include a capacitance sorting function. Simple to operate, the instruments can make quick measurements and simplify the capacitance sorting process. Capacitance is measured by using a small DC voltage signal that slightly charges the connected capacitor for a short time.
protects the meter for better durability. Additionally, the top rubber visor molding above the screen is designed to prevent scratches on the display when the meter is positioned upside down. A single 9V battery or the included DC 12V power adaptor (included with model 830C only) can be used to power the meter. This gives user flexibility for portable or bench-top use.
PACKAGE CONTENTS Each 830C and 890C capacitance meters are shipped with the following contents. • 830C or 890C capacitance meter • Instruction Manual • Mini USB Interface Cable • Red & Black Banana to Alligator Test Leads • 9V Battery • *AC Adapter (model 830C only) *This can be purchased as an optional accessory for model 890C. There are two versions of this adapter available. One version is for 110 V AC line voltage input, and the other for 220 V AC line voltage input.
FRONT PANEL OVERVIEW Figure 1 - Front Panel Display (model 830C shown) 14
Front Panel Display Descriptions 1. Power ON/OFF Button 2. Relative mode / right arrow selection button 3. Input sleeve terminals for positive (+), negative (-), and guard (see “Guard Terminal” in “SUPPLEMENTAL INFORMATION” section for details) 4. Input sockets (banana jack inputs) 5. LCD Display 6. Set HI/LO limit / Back light button 7. Manual and Auto range button 8. Set button for entering HI/LO limit set function menu 9. Tolerance mode / up arrow selection button 10.
WARNING: Before connecting an external power adapter, please check the battery compartment in the rear side of the unit. If a battery is installed, be sure that the polarity matches the (+) and (-) labels as indicated inside the battery compartment. If it is not installed correctly, please remove the battery and install it with correct matching polarity as indicated in the compartment. See “Installing Battery” section for details.
button; meaning that function will be set or configured upon pressing it. To escape out of the functions, hold and press the corresponding button for two seconds. Blue – Some of the buttons have a blue label underneath a white label. This means the function indicated by the blue label will be set if that button is pressed and held for 2 seconds. Yellow – There are total of 6 buttons with yellow labels. They are .
LCD DISPLAY OVERVIEW Figure 2 - LCD Indicator Display LCD Display Descriptions 1. MAX AVG MIN – Static Recording mode, indicates present reading 2. C – Reading indicator 3. AUTO – Auto range 4. REL – Relative mode indicator 5. DH – Data Hold 6.
7. – Beep tone indicator for tolerance and compare mode 8. % - Tolerance percentage indicator 9. kHz – Frequency units indicator 10. – Capacitance units (Farad) indicator (Actual display will show pF, mF, or µF) 11. RMT – Remote mode indicator 12. – Primary display 13. – Low battery indicator 14. – Reading out of the LO limit 15. – Reading out of the HI limit 16. @OFF – Enable Auto Power off indicator 17.
PUSH BUTTONS OVERVIEW Figure 3 - Front Panel Buttons 1. Power ON or OFF 2. Tolerance Mode 3. HI/LO Limits for compare mode and to enable backlight 4.
5. 6. 7. 8. 9.
POWERING INSTRUMENT Before operating the instrument, make sure to connect it to a power source, which could be either a battery or an external AC adapter. Installing Battery The 830C and 890C capacitance meters can be powered by a battery, thus making it portable. The meters use a standard 9V size battery (or NEDA 1604, JIS006P, IEC6F22 carbon-zinc or alkaline battery). To install the battery: 1. Place the meter upside down.
Figure 4 - Back Cover 2. Insert 9V battery into compartment. Note the positive (+) and negative (-) terminals as indicated inside the battery compartment (See Figure 5). Be sure to insert the battery with matching polarity.
Figure 5 - Battery Compartment 3. Place the battery compartment cover piece by sliding it into the top slid first. Place screw at the bottom of the cover piece and tighten down with a screw driver. 4. Push and hold down the button for 2 seconds to turn on the instrument. Connecting External Power Source The 830C and 890C can also be powered using an external AC adapter. The model 830C comes with this adapter included in the package, while it is optional for model 890C. There are two types available.
one that has the correct AC line voltage input for use in your country. For external power, use AC adapter rated for output 12VDC, 150mA, with a standard 4mm connector only. WARNING: Use of incorrect adapters may damage the instrument. Please use B&K Precision’s adapter only. Be sure to use one with the correct AC line input voltage. To connect the adapter, do the following: 1.
2. Connect the AC adapter connector into the right side panel of the instrument. See Figure 6 below. 3. Now, connect the AC Adapter socket into an electrical outlet. 4. Push and hold down the button for 2 seconds to turn on the instrument. 12VDC Input AC Adapter Figure 6 - Connecting AC Adapter to Meter Note: The meter can be operated with a battery installed while an AC adapter is plugged in at the same time (As long as the battery is inserted properly with correct polarity).
Low Battery Indication The capacitance meter has a low battery indicator to notify the user when to replace the battery. When the display starts flashing the indicator, the battery voltage is below normal working voltage. In this case, accuracy of the meter will also decrease. It is recommended that the battery be replaced as soon as possible before continuing operation. See “Installing Battery” for instructions.
When Using Battery Power When the meter is powered using 9V battery, the back light display will turn on upon holding the button for 2 seconds. It will stay at maximum brightness for 15 seconds if backlight settings are set to default value. When Using External Power When the meter is powered using an external AC adapter, the back light display will turn on upon pressing and holding the button for 2 seconds. It will stay at maximum brightness continuously until the user presses and holds the seconds again.
POWER ON OPTIONS To select power on options, press and hold the pushbutton as indicated in the table below while turning on the meter. For example, press and hold and press PUSHBUTTON REL SET to enter setup menu. DESCRIPTION Reset the high and the low limits to factory default values. To see firmware version Setup menu for configuration settings of the meter. Reset HI/LO Limits To reset the high and low limit values to factory default values, press and hold the and press button.
Note: If you released the button before hearing the two beeps, the display will not show any digits on the primary display for a few more seconds. Please wait until it displays digits before operation. View Firmware Version To check the firmware version of the meter, press and hold and press the button. Keep holding the button for two seconds and release. The screen will now show the version. At this point, the power button is disabled.
holding until “bAUd” is displayed on the primary display. For details on configuring settings, please see SETUP OPTIONS section. To exit setup menu, press and hold for 2 seconds and the meter will return to normal display. SETUP OPTIONS There are various system parameters that the meter can be configured for when entering the setup menu (see “Setup Menu” section for instructions on how to access the menu).
ECHO Prnt beep OFF OFF ON LbUt OFF AOFF 15 bLIt 15 bOFF 01 bON 03 dEFA rSt is always 1 bit) Echo: “ON” or “OFF” Print: “ON” or “OFF” Beep: “ON” or “OFF” Lock buttons , “OFF”: enable push buttons “ON”: disable pushbuttons 1~99 minutes, “OFF” means to disable auto power off. 1~99 seconds, “OFF” means to disable turning off backlit automatically.
Baud Rate (bAUd) The baud rate is used for communication with a PC over the USB interface. For PC communication, the baudrate MUST match the settings on the PC side. Selectable rates are (in units of bps): 2400, 4800, 9600, 19200 Note: The secondary display of the meter will show the rates with units as “kHz”, but it’s actually a 3 reference for multiple of 1000 or “x 10 ”. For example, 3 9.6 kHz means 9.6 x 10 = 9600.
Figure 7 - Steps to Change Baud Rates 34
Parity (PArt) The parity setting is used for communication with the PC over USB interface. For PC communication, parity MUST match the settings on the PC side.
Data Bits (Data) The data bits setting is used for communication with a PC over USB interface. For PC communication, data bits MUST match settings on the PC side.
Echo (ECHO) The echo function will set the meter to “echo” or return all characters that are received. For example, if “*IDN?” is sent to the meter, it will return “*IDN?” as well as the queried information.
Print Only (Prnt) Under remote operation, when the meter is set with print only mode to “ON”, it will print out measured data when the measurement cycle is complete and automatically send the newest data to the PC continuously. In this mode, any commands sent from the PC will not be accepted. The remote indicator on the meter display will flash when this mode is enabled.
Figure 11 - Steps to Change Print-Only Setting Key Beeper (bEEP) The key press beep sound can be enabled or disabled.
Figure 12 - Steps to Toggle Key Keep Lock Buttons (LbUt) The front push buttons can be locked by changing the settings of this option. When “ON”, buttons will be locked and cannot be pressed upon turning on the meter. All buttons are disabled at this point except the button, which can be used to turn off the meter. The power-on option can still be operated even if buttons are locked.
with a means to disable or unlock the buttons when necessary. Selectable parameters: OFF, ON Figure 13 - Steps to Change Lock Buttons Settings Auto Power Off (AOFF) This option will allow a timer to be set for auto power off.
indicator. When disabled, the meter will run indefinitely. Note: The timer will not be in effect if any of the below happens: 1. Push buttons are used. 2. Static recording is set. 3. Auto power off is disabled or “OFF”. If any of the above conditions do not apply, the timer will start counting based on the timer setting configured.
Figure 14 - Steps to Change Auto Power Off Settings 43
Backlit Display Time (bLIt) Back light of the meters can be turned off automatically with a timer. Once set, back light will remain on for the period of the timer setting upon enabling the back light by pressing and holding for two seconds. If it is set to “OFF”, the back light will remain ON indefinitely. Selectable parameters: OFF, 1 – 99 seconds Note: To conserve and extend battery life, select a shorter timer setting.
Selectable parameters, OFF, 01, 02, 03 Note: “03” is the brightest and “01” is the dimmest.
Figure 15 - Steps to Change Back light Off State 46
Brightness Level On State (bON) This option allows user to set the brightness level for the on state of the back light. When it is set to “OFF”, back light will be in the lowest brightness level in the on state. If a level is set, the backlight will be much brighter when backlight is enabled by pressing the button. Selectable parameters, OFF, 01, 02, 03 Note: “03” is the brightest and “01” is the dimmest level. However, when set, “OFF” state will still be the lowest brightness level.
Figure 16 - Steps to Change Back light ON State 48
Reset to Default (dEFA) This option will reset all setup option parameters back to default values as listed in Table 1 - Setup Menu System Settings and Default. To reset, simply press and hold the button for two seconds. The meter will then be resetting all parameters immediately. All push buttons will not respond immediately, so wait for approximately 5 seconds until the display changes and returns to “bAUd” option.
Figure 17 - Reset to Factory Default 50
OPERATION INSTRUCTIONS Data Hold/Trigger Hold The data hold/trigger hold function allows the user to freeze the display when pressed, holding the measured value until data hold is turned off. Turn On Data Hold To use data hold, press the button once. The “DH” indicator will display on the screen when data hold is active. Trigger Hold Users can send a trigger hold once data hold has been turned on.
Figure 18 - Data/Trigger Hold Operation 52
Static Recording This mode is used for recording maximum, minimum, and average capacitance values. It captures only stable values and updates to memory. Values that result in displaying “OL” or that are below 10 counts are not recorded. This mode is useful for determining the range of values a component falls into. Enable Static Recording Press the button to enter the static recording mode. The display should indicate “MAX AVG MIN” simultaneously.
Recording Mode Maximum Mode Mode Average Mode Minimum Recording Mode This is the default mode when first enabling static recording. In this mode, the “MAX AVG MIN” indicator will be displayed. At this point, the meter will start recording measured values applied to the input sockets or terminals. During the recording process, maximum, minimum, and average values will be stored. A beep will sound once a recording has been stored.
Minimum Mode In this mode, the “MIN” indicator will be displayed. This indicates that the value in the primary display represents the recorded minimum value. Average Mode In this mode, the “AVG” indicator will be displayed. This indicates that the value in the primary display represents the recorded average value. This average value is obtained by taking the maximum and minimum recorded values and taking the average of the two values.
Figure 19 - Steps to Use Static Recording 56
Relative Mode Relative mode is used when the user wants to “zero” the meter based on a reference value or wants to obtain a reading that is relative to a reference value. For example, if test leads are used in the measurement, the user may want to set a reference with the test leads inserted into the input terminals so that any measurements taken will not take into account the test leads. Setup Relative Mode To setup the relative mode, simply press the button once.
have test leads or wires connected to the meter in the fashion that they desire. Then, press the button once to “zero” out the meter as reference. This way, any measurements taken would not be affected by the test leads or setup. Note: Relative mode can be set in both auto and manual range, but cannot be set when “OL”, or overload is displayed on the meter. Disable Relative mode To disable relative mode, simply press and hold the button for two seconds.
Figure 20 - Steps to Use Relative Mode 59
Tolerance The tolerance mode feature is specifically used for component sorting purposes. Users who need to test and sort through a large number of components will find this function quite helpful. Tolerance Range The tolerance function is configured primarily by percentage range, meaning a percentage is used to define whether a measured value is within tolerance or out of tolerance. Selectable tolerance range is: 1%, 5%, 10%, and 20%. Setup Tolerance 1.
capacitance is connected to either the input sockets or terminals. WARNING: Be sure that the capacitor is fully discharged BEFORE inserting it into the input sockets or terminals. For large capacitors, it may take longer periods of time for a full discharge. Inserting a charged or partially charged capacitor into the meter’s input sockets or terminals may produce an electric hazard and may also damage the instrument, thus rendering it unusable.
2. Once the desired measured reading is displayed, press the button once to store the reading as the “standard” value or test reference value. At this point, the “TOL” will be displayed on the screen, indicating that the tolerance mode is activated. 3. To select the tolerance range, press the button once more. For each button press, the meter will cycle through the tolerance range percentage in this order: 1%, 5%, 10%, 20%.
Note: Under the following instances, the tolerance mode cannot be set: 1. After setting recording mode 2. After setting HI/LO limit for compare mode 3. The display shows “OL” for overload. Disable Tolerance Mode To disable or exit out of tolerance mode, simply press and hold down the button for two seconds. The “TOL” and/or the percentage indicators “1%”, “5%”, “10%”, or “20%” will disappear on the LCD display.
Figure 22 - Steps to Setup Tolerance Mode 64
Compare Mode This function allows you to sort capacitors and to configure 25 sets of limit ranges. The meter comes with preconfigured stored values for high and low limits.
17 18 19 20 21 22 23 24 25 2200 2700 3300 3900 4700 5600 6800 8200 10000 1980 2430 2970 3510 4230 5040 6120 7380 9000 Setting Compare Mode The following procedure demonstrates how to set compare mode. 1. Press button to enter HI/LO audible alert compare mode. The range will be locked in this mode. The meter will display the “ ” indicator on the upper right. The secondary display will indicate “C01” to “C25” depending on which set was last saved during previous operation.
2. Press the button to cycle through the sets from “C01” to “C25”. 3. Press the button to see the HI and LO values used for the comparison. As an example, the display will cycle through set 5 in this order: “C05”, “L05”, “H05”. The values will briefly display for one second and will return to normal compare display. In the example, it would just return to display “C05” on the secondary display. 18.
Exit Compare Mode To exit compare mode, simply press and hold for two seconds. The “ ” indicator will disappear from the upper right corner of the display, indicating that meter is no longer in compare mode.
Figure 23 - Steps to Setup Compare Mode 69
Figure 24 - Steps to Sort by Compare Mode 70
Setting HI and LO Limits The yellow keys from the front key buttons are used for setting HI and LO limits for compare testing. The button enters the HI/LO limit setup menu. The arrow buttons , , , and are used to adjust the HI/LO sets and HI/LO limit values. The adjusted limit values. button is used to save the Follow the below procedure to configure HI/LO limits. 1. Press and hold for two seconds to enter HI/LO limit setting mode. 2.
3. Once the digit you want to adjust is blinking, use the or to increase or decrease the value of the blinking digit. 4. Once the desired limit value is set, press and hold button for two seconds or until all the digits on the primary display stops blinking. 5. To set the “HI” limit value for the first set, press the button once and the secondary display will show “H01” blinking slowly. 6. To adjust the next set of limit values, simply press the button once.
Exit HI/LO Limit Menu Once all limit values are configured for all the sets to be used, hold and press the seconds to exit.
Figure 25 - Steps to Set HI/LO Limits 74
REMOTE COMMUNICATION The meter has the capability to communicate with a PC over the mini USB interface. Upon installation of a USB driver, the PC can control the instrument over virtual COM. The mini USB communication interface of the meter is fully duplex and has a 64-byte input and output buffer, making it reliable and efficient for data transmission. Connecting Instrument to PC Follow the below procedures for connection setup. 1. Download the USB drivers from www.bkprecision.com . 2.
port. Windows will automatically assign a COM port to the instrument. Please verify which COM port Windows has assigned by going into “Device Manager”. USB (Virtual COM) Configuration The USB will be recognized as a virtual COM on the PC, thus serial port settings must be configured properly for remote communication to be successful. Below are the default settings used by the 830C and 890C meters. They can be changed within the setup menu. See SETUP OPTIONS section for details.
Command Protocols Overview of Command Type and Format All commands are entered in either the upper case or the lower case. There are two types of the meter programming commands: IEEE 488 common commands and Standard Commands for Programmable Instruments (SCPI). Some commands are device-specific to the meter. They are not included in the version 1999.0 of the SCPI standard. However, these commands are designed with the SCPI format in mind and they follow the syntax rules of the standard.
following example shows a command and its associated query: CONFigure:CAPacitance 10u Set the primary display to the capacitance measurement, and select the 10 uF range. CONFigure? Return the function of the primary display measurement. CONFigure is a root level keyword with the second level keyword, CAPacitance, and 10u is the command parameter. The query command ends with a question mark “?”. Note: SCPI stems from IEEE488.1 and IEEE 488.2. Although the IEEE 488.
Responding Message Returned result After the meter executes a query command, the return of the result will be in the following format: + Data Types Returned message is an ASCII string from the meter responding to a query. A query is a command accompanied a “?” mark. Table 3 below explains the different data types.
or “1” for Boolean query command A string is used as command parameters with short literal form HOLD SCPI Commands This section described all the SCPI commands supported by the meter.
*IDN? Description: Queries the instrument ID. Response: , , LLO Local Lockout. This means that all front panel buttons are disabled except for button. GTL Go to local. Puts the meter into the local state, clearing the remote state and front panel lockout. CONFigure Subsystem CONFigure:CAPacitance [] Description: Set capacitance range. Parameter: : Measuring unit is Farad.
CONFigure? Description: Query the function of the primary display. Response: Return <”function range, resolution”> format string, the examples are as following: CAP +1.000000E07,+1.000000E-1 Capacitance 100nF 0.01nF CALCulate Subsystem: CALCulate:FUNCtion NULL TOLerance LIMit AVERage NONE Description: Enable or disable the function of the calculation.
internal trigger source. Example: CALC:FUNC AVER Note: The trigger source is changed to immediate when the AVERage function is enabled. User can combine TOL/NULL, LIM/NULL, AVER/NULL or NULL/AVER functions, but the other combinations are void. CALCulate:FUNCtion? Description: Query the function of the calculation. Response: Return format string: AVER, TOL, LIM, NULL or NONE.
CALCulate:NULL:OFFSet? Description: Query the offset value of the null (relative) function. Response: Return format string. Note: If relative function is disabled, querying the offset of relative mode will return an error. Reenable in order to get a returned value. Use CALCulate:NULL:STATe command shown below to control the relative function. CALCulate:NULL:STATe Description: Enable or disable the null (relative) function of the calculation.
Example: CALC:AVER:MAX? ; Query the maximum recorded value. CALC:AVER:MIN? ; Query the minimum recorded value. CALC:AVER:AVER? ; Query the average recorded value. CALC:AVER:PRES? ; Query the present recorded value. Note: If static recording function is disabled, all queries will return an error message on display until the function is re-enabled again. Use CALCulate:AVERage:STATe command shown below to control static recording function.
Description: Limit function. Set the mode or the value of the Parameter: : The value of mode is from 01 to 25. : The limit value is from 00000 to 11000. Example: CALC:LIM:MODE 01 ; Apply the settings of the mode 1. CALC:LIM:UPP 07,01000 ; Set the upper limit th value of the 7 set. CALCulate:LIMit :MODE? :LOWer? :UPPer? Description: Query the mode or the value of the Limit function. Parameter: from 01 to 25. : The value of mode is Response: Return format string.
Limit function. CALC:LIM:UPP? 07 of the mode 7. ; Query the upper value Note: CALCulate:LIMit:MODE? Query will respond with an error if meter is not in limit function and if the state of limit function is disabled. To set meter to limit function, use the CALCulate:FUNCtion command. CALCulate:LIMit:STATe Description: Enable or disable the Limit function of the calculation.
CALCulate:TOLerance:RANGe command will return an error on display. CALCulate:TOLerance:BASE? Description: Query the base value of the tolerance function. Response: Return format string. CALCulate:TOLerance:VALue? Description: Query the tolerance value of the tolerance function. Response: Return format string. The returned value will not be in terms of %. For example, if display shows 0.35%, return value could be 3.
TRIGger:Subsystem: TRIGger:SOURce < BUS | IMMediate > Description: Select a source of the trigger signal. Parameter: BUS – select a bus command source and enter the trigger (hold) mode. In this mode, data hold will be activated to hold each measured reading after trigger (*TRG) is received from the bus. IMM – select the internal trigger source. Example: TRIG:SOUR BUS Note: The AVERage function will be disabled when entering into trigger mode.
Example: *TRG Response: Return display value, e.g +1.23450000E+00 Measuring Subsystem: FETCh? Description: Response: Return the primary display value of device’s output buffer. Retrieves the measurements taken by the INITiate command and places them into the device’s output buffer. The query can return data any time that the last reading is valid. The meter generates an error -230 and no result is returned when the meter is in the trigger idle state and the data is invalid. Return format string.
SYSTem Subsystem SYSTem:ERRor? Description: Query the error message. Response: Return format string. Table 4 is a list of SCPI error message that might occur during operation. Note: The buffer size of the system error queue is one. Keep the last error.
Description: Query the version of the SCPI. Response: Return 1999.0 string. SYSTem:BEEP Description: Enable or disable the key beep. Parameter: : ON; OFF. SYSTem:BLIT Description: Enable or disable the backlit. Parameter: Note: : ON; OFF. The backlit is turned off automatically when the backlit time isn’t zero and the time comes up. SYSTem:BLIT:TIME Description: Set a backlit time. Parameter: : 0 ~ 99.
SYSTem:AOFF:TIME Description: Set an auto power off time. Parameter: : 0 ~ 99. Set “0” to disable auto power off function. Note: Stored in non-volatile memory. SYSTem:DEFAult Description: Setting default value to the nonvolatile memory includes SETUP value but don’t affect the other calibrated data. STATus Subsystem: STATus? Description: Query the status of the meter. Response: Returns a <”ABCDEFGHIJKLMNOPQRSTU”> format string (21 characters).
Each character represents the following: Item Explanation Average 0:off, 1:on A function B Null function 0:off, 1:on Limit 0:off, A:mode 1 ~ C function Y:mode 25 Limit result 0:none, 1:Go, 2:Lo D nGo, 3:Hi nGo Tolerance 0:off, 1:1%, 2:5%, E function 3:10%, 4:20%, F:no range setting Tolerance 0:none, 1:pass, F Result 2:fail Trigger I:IMM, B:BUS G source H Unused always 0 I Unused always 0 J Beep 0:off, 1:on Auto Power 0:off, 1:on K Off L Back lit 0:off, 1:on M Meter mode L:normal N Unused always 0 O Unused a
S T U Battery Unused Auto range 0:normal, 1:low always 0 0:off, 1:on Summary of SCPI Commands Command CONFigure? CONFigure :CAPacitance Parameter [] CALCulate :FUNCtion? :FUNCtion :NULL :OFFSet? :STATe :AVERage :MAXimum? :MINimum? :AVERage? :PRESent? :STATe Explanation Query display function Configure the meter perform specified measurement Query the calculation function. Set a calculation function.
:LIMit :MODE :UPPer :LOWer :MODE? :UPPer? :LOWer? :STATe :TOLerance :RANGe , , Query the value of the limit function Enable/disable the limit function <1 | 5 | 10 | 20> Set the value of the Tolerance function Query the value of the Tolerance base Query the tolerance value Enable/disable the tolerance function :BASE? :VALue? :STATe Set the value of the limit function TRIGger :SOURce? :SOURce FETCh? SYSTem :AOFF:TIME
:DEFAult Set default value to the non-volatile memory Query error message Query version of the SCPI Query the status :ERRor? :VERSion? STATus? Error Codes In certain situations, errors may occur, and an error code will be displayed on the meter. Below defines the error description based on the error code.
measures capacitance by charging the capacitor with a known current, measuring the resulting time of the charging period, and then calculating the capacitance. The larger capacitors will take longer time to charge. Some large capacitors can take over a minute to measure due to the long charge time required. A flashing “C” indicates charging capacitor.
Measuring large capacitors: Measuring very large capacitors, for example a 100 mF value, requires a long time. . In this case, after inserting the capacitor into the meter’s terminals, the meter display may seem unresponsive and not respond to key presses. This is normal, and indicates that the meter is currently charging the capacitor. Depending on the size of the capacitor, this charging period can take up to 2 minutes until the meter becomes responsive and display the result.
The guard terminal is used to minimize noise and to help minimize parasitic components of the capacitor to be measured, thus allowing for more accurate results. SPECIFICATIONS Below are some remarks in regards to all specifications pertaining to both 830C and 890C capacitance meters. *Specifications are subject to change without notice. The following conditions must be met for the specifications to be valid: 1. Measurements are performed at the test socket. 2.
General specification Display: 5 digit liquid crystal display (LCD) with maximum reading of 11000 counts. Automatic polarity indication. Function: Capacitance measurement by DC charge and discharge method. Visible and audible Tolerance mode assists you to sort capacitors Min/Max/Average, Data Hold with Manual or Auto Trigger and Relative modes Comparison mode with 25 sets of HI/LO limits can be selected Backlit display for easy reading in the dark.
The " " appears when the battery voltage drops below 6V (approx.). Operating temperature: 0°C to 40 °C (32ºF~113ºF) Storage temperature: -20°C to 50°C (-4ºF~122ºF) without battery Relative Humidity (R.H.) : Maximum 80% R.H. for temperature up to 31°C decreasing linearly to 50% R.H. at 40°C Temperature coefficient 0.
Safety Standard: Designed in compliance with EN61010-1 (IEC1010-1) for CAT-II 50V, Pollution Degree II Environment. EMC designed in compliance with EN61326-1. Electrical specifications Accuracy is given as (% of reading + counts of least significant digit) at 23°C ± 5°C, with relative humidity less than 80% R.H. Capacitance Range Resolution Accuracy* Measuring rate as full scale (approx.) 5 times /sec 5 times /sec 5 times /sec 5 times /sec 5 times /sec 5 times /sec 0.86 times /sec 0.13 times /sec 1000.
Specifications are subject to change without notice. To ensure the most current version of this manual, please download the current version here: http://www.bkprecision.com/search/manual/830C For current up-to-date product information, please visit www.bkprecision.
MAINTENANCE WARNING: Do not perform any service by yourself. Service should only be done by qualified personnel and trained technicians. Service If the instrument fails to operate, check battery and test leads. Replace them as necessary. If the instrument still cannot work, verify with the operating instructions to make sure correct procedures are followed. When servicing, use specified replacement parts only. Note: The meter must be completely turned off while replacing the battery.
● Push buttons failure ● ● ● Failed on Remote control ● ● Display is frozen and buttons do not respond Check whether the meter is under remote control. Turn OFF and then turn ON the meter again. Check the setup mode whether “Lbut” of lock buttons has been set “on”. Check the baud rate, parity, Data bit, Stop bit (default is 9600, n, 8, 1) Check USB driver installation.
Before cleaning this meter, make sure the power is OFF and remove external AC adapter if one is used. To clean the meter, wipe the dirty parts with a soft cloth soaked with diluted neutral detergent. Avoid having the instrument too wet to prevent the detergent from penetrating into the inside components of the meter, causing damages. After cleaning, make sure the instrument is completely dried before operating it again.
SERVICE INFORMATION Warranty Service: Please go the support and service section on our website www.bkprecision.com to obtain a RMA #. Return the product in the original packaging with proof of purchase to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Non-Warranty Service: Please go the support and service section on our website www.bkprecision.com to obtain a RMA #.
LIMITED WARRANTY B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of three years from date of purchase. B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt.
INDEX Auto Power Off (AOFF), 41 Backlit, 27 Battery, 22 Baud Rate (bAUd), 33 Compare Mode, 65 Data Bits (Data), 36 Data Hold, 51 dEFA, 49 Echo (ECHO), 37 Error Codes, 97 External Power, 24 Firmware, 30 Guard Terminal, 99 HI and LO Limits, 71 Key Beeper (bEEP), 39 LCD Display, 18 Lock Buttons (LbUt), 40 Low Battery, 27 Parity (PArt), 35 Print Only(Prnt), 38 Relative Mode, 57 Remote Mode, 76 SCPI, 77 Setup Menu, 30 Static Recording, 53 Tolerance, 60 USB drivers, 75
22820 Savi Ranch Parkway Yorba Linda, CA 92887 www.bkprecision.com © 2011 B&K Precision Corp.