User Manual
60 
6.  To disable list operation, first press   to disable the input and then press   and  
(or press  ). Select Off and press  to confirm. The Trig annunciator will 
disappear when list operation is disabled. 
3.8  Battery Test Function 
The load has a built-in battery test function that uses CC mode and calculates the battery 
capacity using a fixed current load. The test’s stop conditions can be specified by the following:  
Stop Voltage: Set a cut-off voltage level. When this level is reached, the test will end. Valid 
range: 0 V – max. rated voltage. 
Stop Capacity: Sets a capacity level. When the calculated capacity reaches this level, the test 
will end. Valid range: 0 Ah – 999.99 Ah 
Stop Timer: Sets a timer. When the specified time is reached, the test will end. Valid range: 0 
s – 99999 s. 
During operation, the load will sink current at the specified level while measuring and updating 
the capacity value in amp-hours (Ah) and tracking the time period of the test. When any one of 
the three stop conditions above is met, the test will end, and the display will show the 
measured capacity and time results. 
To setup and run the battery test function, follow the steps below: 
1.  Disable the input of the load and connect it to the battery to be tested. 
2.  Press   and use the numeric keypad or rotary knob to enter a current sink value 
for the test. Press  . 
3.  Press  ( ) and  ( ) to configure and enable the battery test function.  
4.  The load will prompt to enter all three test stop conditions, starting with Stop Voltage 
value. Use the numeric keypad or rotary knob to enter a voltage level. Then press 
to continue. 
5.  The load will prompt to enter the Stop Capacity. Use the numeric keypad or rotary 
knob to enter a capacity value in Ah. Then press  to continue. 