3.  Enter a value for Voltage On Delay. This can be set between 0.00s and 99.99s. Then 
press   to continue. 
4.  Follow the same steps for Current Range. The value must be within the maximum 
input limits of the load. Press   to continue. 
5.  Set the Start Current. The value must be less than Current Range. Press  to 
continue. 
6.  Do the same for Step Current and End Current. Press  after each setting to 
continue. 
7.  Enter a value for OCP Voltage and press  to continue. 
8.  Finally, enter a value for Max Trip Current, and then Min Trip Current. Press   
after each settings to continue. 
9.  The load will prompt to Save OCP File. Select a number between 1 and 5. Then press 
 to save all settings to the selected location. 
Note: At any time when configuring any parameters for the test, you can press the   
key to select the previous parameter to edit. 
To recall the settings: 
1.  Press  ( ) and  to enter the OCP Test menu. Select Recall and 
press  . 
2.  Use the keypad or knob to select the saved location number and press  . 
All settings will be recalled when selecting Edit from the OCP Test menu. 
To run the OCP test: 
1.  Recall the settings from memory by following the steps above. 
2.  Then, from the OCP Test menu, select Run and press  . The display 
will show the following: 