User Manual
72 
Start Current 
This is the starting current value when OCP test begins. The settable 
range of this value is limited by Current Range. 
Step Current 
This is the size of the step current as the test runs from the Start Current 
to End Current. The test will only run through in steps if the measured 
input voltage is greater than the OCP Voltage. Otherwise, the test will 
end after running the Start Current for the amount of time specified in 
Step Delay. 
Step Delay 
This is the delay time to hold each current step in the test. This 
determines how fast or slow to run through the test. 
End Current 
This is the ending current value before OCP test ends. The test will only 
run through in steps from Start Current to the End Current if input 
voltage is greater than the OCP Voltage.  The settable range of this value 
is limited by Current Range. 
OCP Voltage 
This value is a limit that is used to test for OCP conditions. In the event of 
an over current condition, when the input voltage is greater than this 
value, the test will run through from Start Current to End Current in 
steps. The input current (the over current) value will be checked during 
this process. If current is within the Max Trip Current and Min Trip 
Current boundaries, the load will indicate Pass at the end of the test. If 
outside of these boundaries, the load will indicate Fault. 
Max Trip Current 
The maximum acceptable current limit of the over current value. 
Min Trip Current 
The minimum acceptable current limit of the over current value. 
The following is a step by step procedure to configure an OCP test program. 
1.  Press  ( ) and  to enter the OCP Test menu. The following will be 
displayed: 
2.  Select Edit and press  . Enter a value using the numeric keypad or rotary knob 
for Voltage On Level. The value must be within the maximum input limits of the load. 
Press   to continue. 
OCP   TEST 
Run      Recall        Edit 