User Manual
67 
0 ≤ T
pf 
≤ (T
on
+ T 
off
) 
Tpf = P/F Delay Time 
T
on
 = On Time 
T
off 
= Off Time 
Use the numeric keypad or rotary knob to enter a value. Then press  . 
8.  If there are more than one active sequence, the load will repeat the prompts to 
configure the On Time, Off Time, and P/F Delay Time. Follow steps 7 – 9 for each 
sequence. 
9.  Once all sequences’ parameters are configured, the display will show: 
  Select Complete-Stop or Failure-Stop, then press  . 
  Complete-Stop: Stop test when program is complete. 
  Failure-Stop: Stop test when testing fails. 
10. The load will prompt to enter a value for Chain Program File. This specifies the next 
program to run immediately after the end of the test. Select between 0 – 10. Select 0 if 
no other programs will be run at the end of the test. Otherwise, select the program 
number to run after the test is completed. Use the numeric keypad or rotary knob to 
enter a value and then press  to confirm. 
11. Lastly, the load will prompt to select a number to Save Program File. This is the 
program number that will be referred to when recalling all the configured settings from 
previous steps or chaining with another program. Select between 1 – 10 with the 
PROGRAM 
Complete–Stop      Failure–Stop  