7.  Press   and it will prompt to enter the slew rate of the step. If High-Rate was 
selected earlier, units will be in A/us. Otherwise, it will be in A/ms. 
8.  Press  and it will prompt to enter the step with. Valid range is 20 us – 3600 s. 
Enter a value and then press  to continue. 
9.  The load will prompt to enter parameters for the next step. Follow steps 6-8 to enter 
the values for the step’s parameters. Once all steps’ parameters are entered, the load 
will prompt for the Repeat Count, which sets how many times to repeat the list 
program. Valid range is 1 – 65535. Press  to continue. 
10. The load will prompt to select the memory location to store the list parameter 
information. There are seven total save locations. Enter a value between 1 and 7, then 
press   to confirm. The display will return to the List menu. 
Note: If List mode is enabled, the Edit menu will be locked and list parameters cannot be 
changed until List mode is disabled. 
EDIT  LIST 
Step 001    Width = 0.00002S  
EDIT  LIST 
Step 001    Rate = 0.0001A/us  
EDIT  LIST 
Step 001    Level = 0.0000A  