User Manual
52 
Figure 20 - Continuous Transient Operation Current Waveform 
Pulse 
In this mode, the electronic load generates a transient pulse of programmable width when 
pulse transient operation is in effect. 
In pulse mode, you can set A/B level, the pulse width, and A/B slew rate. The electronic load 
will automatically switch to A level after maintaining A width time. Then it will switch to B level. 
The electronic load will not switch to A level again until the instrument receives a trigger signal. 
The following figure shows the current waveform in pulse transient operation. 
Figure 21 - Pulse Transient Operation 
Toggle 
In toggle mode, the electronic load will switch between A level and B level when receiving a 
trigger signal. The following picture shows the current waveform in toggle transient operation. 