User Manual
25 
The setup parameters are: Range, High (Voltage limit), and Low (Voltage limit). Use the 
( or  ) key to select each parameter, and use the numeric keypad to change the 
value. Press  to confirm the change. 
Range 
Use the numeric keypad to change the range. This value will also act as a limit to how much 
resistance the load can be configured to. The adjustable ranges vary depending on the model. 
See specifications for details. 
High 
This parameter refers to the voltage high limit for the automatic test mode. During automatic 
test mode, the device under test (DUT) must be operating below the configured value for the 
test to PASS upon completion. If the DUT operates above the configured value, the test will FAIL 
upon completion. 
Note: This parameter is used for Automatic Test Function ONLY. 
Low 
This parameter refers to the voltage low limit for the automatic test mode. During automatic 
test mode, the DUT must be operating above the configured value for the test to PASS upon 
completion. If the DUT operates below the configured value, the test will FAIL upon completion. 
Note: This parameter is used for Automatic Test Function ONLY. 
Constant Power (CW) Mode 
In this mode, the electronic load will consume a constant power. When input voltage increases, 
the input current will decrease, while power (P = V*I) will remain the same. 
Constant Resistance 
Range=7500.0Ω 