User Manual

84
Models
8610
8612
8614
8616
Transient Mode (CC mode)
T1 & T2
1
20 µs 3600 s/Resolution: 10 µs
Accuracy
5 µs + 100 ppm
Slew Rate
2
Low
Rate
0.001-0.25 A/µs
0.0001-0.1 A/µs
0.001-0.25 A/µs
0.0001-0.1 A/µs
High
Rate
0.01-2.5 A/µs
0.001-1 A/µs
0.01-2.5 A/µs
0.001-1 A/µs
Accuracy
within 40% of programmed value
Measurement
Readback Voltage
Range
Low
0 18 V
0 50 V
0 18 V
0 50 V
High
0 120 V
0 500 V
0 120 V
0 500 V
Resolution
Low
0.1 mV
1 mV
0.1 mV
1 mV
High
1 mV
10 mV
1 mV
10 mV
Accuracy
±(0.05%+0.05%FS)
Readback Current
Range
Low
0 12 A
0 3 A
0 24 A
0 6 A
High
0 120 A
0 30 A
0 240 A
0 60 A
Resolution
Low
1 mA
0.1 mA
1 mA
0.1 mA
High
10 mA
1 mA
10 mA
1 mA
Accuracy
±(0.05%+0.1%FS)
±(0.05%+0.05%FS)
±(0.05%+0.1%FS)
±(0.05%+0.05%FS)
Readback Power
Range
750 W
1500 W
1200 W
Resolution
10 mW
100 mW
Accuracy
±(0.2%+0.2%FS)
Protection Range (typical)
OPP
760 W
1550 W
1250 W
OCP
Low
13.2 A
3.3 A
26.4 A
6.6 A
High
132 A
33 A
264 A
66 A
OVP
130 V
530 V
130 V
530 V
OTP
85
General (typical)
Short Circuit
Current
(CC)
Low
13.2 A
3.3 A
26.4 A
6.6 A
High
132 A
33 A
264 A
66 A
Voltage
0 V
Resistance
10 mΩ
120 mΩ
6 mΩ
60 mΩ
Input Terminal
Impedance
300 kΩ
1 MΩ
300 kΩ
1 MΩ
Notes:
1) Fast pulse trains with large transitions may not be achievable.