3.  Press   to start the test. When the test is running, Run will be indicated 
in place of Stop. When the test ends, either Pass or Fault will appear next 
to Stop. 
4.  To stop the test at any time, press  . 
OPP Test Function 
The load has a built-in OPP test function that can be used to automate testing for OPP 
conditions. The load can save up to 5 OPP test programs into internal memory, which can be 
recalled and run. 
The OPP Test program requires the following parameters be configured before it can be run. 
Voltage On Level 
This is the minimum voltage level the load measures before the start of 
the test. 
Voltage On Delay 
This is the delay time from when the Voltage On Level is reached to the 
actual start of the test. 
Current Range 
This specifies the current range of the test. All models have two current 
ranges, with the lower range allowing for higher resolution. This value 
will also limit the maximum settings for other parameters configured for 
the test program. 
Start Power 
This is the starting power value when OPP test begins. 
Step Power 
This is the size of the step power as the test runs from the Start Power to 
End Power. The test will only run through in steps if the measured input 
voltage is greater than the OPP Voltage. Otherwise, the test will end 
after running the Start Power for the amount of time specified in Step 