User Manual
71 
If there is a failure, fail will be indicated in place of pass. At this point, you can press 
( ) to view which sequences were tested and passed/failed. It will be 
indicated like the following: 
Y indicates the tested sequence(s) that passed/failed. If a sequence has been skipped in 
the test, its sequence number will be indicated instead of a Y. 
7.  To stop a running program at any time, press   ( )once. To completely exit 
the automatic test function, press   a few times until the load returns back to the 
normal display. 
OCP Test Function 
The load has a built-in OCP test function that can be used to automate testing for OCP 
conditions. The load can save up to 5 OCP test programs into internal memory, which can be 
recalled and run. 
The OCP Test program requires the following parameters be configured before it can be run. 
Voltage On Level 
This is the minimum voltage level the load measures before the start of 
the test. 
Voltage On Delay 
This is the delay time from when the Voltage On Level is reached to the 
actual start of the test. 
Current Range 
This specifies the current range of the test. All models have two current 
ranges, with the lower range allowing for higher resolution. This value 
will also limit the maximum current settings for other parameters 
configured for the test program. 
10.000V 
0.000A 
PRG01:   □□□□□□4Y2Y  pass 
OFF CC 