User Manual
57 
Configure List 
Follow the steps below to configure list operation: 
1.  From the front panel, press   and then   (or press  ). The following 
display will show: 
2.  Select Edit and press  . The following display will show: 
3.  Select either High-Rate or Low-Rate. These options configure the settable slew rate 
range. If High-Rate is selected, users can adjust the slew rate of the transient in the 
A/us range. If Low-Rate is selected, users can adjust the same in A/ms range. The 
settable range will vary depending on the model. 
4.  The load will then prompt to enter a value for Current Range. Use the numeric keypad 
or rotary knob to set the value for current range, then press  to confirm. 
5.  The load will then prompt to enter a value for File Step, which is the total number of 
steps for the list. Use the numeric keypad or rotary knob to set the value, then press 
to confirm. Valid range is 2 – 84. 
6.  The load will prompt to enter values for each steps parameters, starting with the step 
level. Use the numeric keypad or rotary knob to enter a value. 
EDIT  LIST 
High-Rate      Low-Rate 
LIST 
On      Recall       Edit 