User Manual
16 
Table 2 - Self-test Messages 
Error Message on Display 
Description 
EEPROM FAILURE 
The internal EEPROM is corrupted or damaged. 
Config Data Lost 
The last operation data within the EEPROM is lost. 
Calibration Data Lost 
Calibration data within the EEPROM is lost. 
FactoryCal.Data Lost 
Factory calibration data is lost. 
MainframeInitialize Lost 
The system settings within the EEPROM is lost. 
If any of these errors occur, please contact B&K Precision. 
Input Check 
Follow the steps below to check that the load is operating correctly. A DC power supply rated 
for at least 5V and 1 A will be required to proceed with this check. 
1.  Power on the load.  The display will show the OFF annunciator above the voltage display. 
2.  Connect the input terminal to a DC power supply and configure the supply to output 5 V 
and current limit to 1 A. 
3.  Turn on the DC power supply’s output. Observe the load’s measured voltage display, 
which should read close to or exactly 5.000V. 
4.  Press   so that its back light turns on, and enter 0.500 A using the numeric 
keypad. Press  . 
5.  The display should show CC = 0.500A on the bottom right. 
6.  Press  and its backlight will be lit. The OFF indicator will change to CC and the 
measured current should now display a value close to 0.500 A. 
7.  This setup verifies that the load is drawing power correctly from the power supply. 
Note: If the load is not drawing power from the DC power supply, check all load 
protection limits and settings within the menu to verify that the load is configured 
to allow drawing power at 5V, 0.500 A. Also, verify that the CC mode parameters 
are setup to operate within the configured valid ranges by pressing  (  ) 
and  .  