Setup List Parameters
Below is an example with step by step instructions on setting up a sequence using
list modes.
Step 1 – Select step number to
program
If not already, boot the instrument into
LIST mode by holding the key
while switching on the power button.
Press , then to enter
the LIST menu for programming LIST
parameters. The display will look like
the one on the right.
If you want to program from the first
step, leave it shown on the bottom
display as 0. In this example, we will
start from 0. After entering all
parameters assigned for the step, the
menu system will automatically go
back to this setup screen with step
number automatically incremented by
1. You only need to adjust this if you
want to edit a programmed step later
on or if you want to program a
different step.
Press to go and configure
the next parameter.
ShiftShift
BB
View
C
Recall
CC
Short CV
OFF
Shift
30A
CR
A
B
C
D
Set
V
W
A
150W DC Electronic Load
8540
Step 2 – Select type
Bottom display will show the type for
step 0. Use the knob to select
between CC, CV, or CR type. When
running the list, this step will set the
load to the selected type of operation.
In this example, we will select CC.
Press to go and configure
the next parameter.
Recall
CC
Short CV
OFF
Shift
30A
CR
A
B
C
D
Set
V
W
A
150W DC Electronic Load
8540
View
C
View
C
Programmable Range: 0 – 99
Use the keys to change cursor
position shown by the LED light next to
either A, B, C, or D (These four denotes
cursor position, with A selecting the most
significant figure and D the least significant
figure.
BB
AA
NOTE
Users can create a sequence with a
mixed combination of different types in
each step (i.e. One step can run in CC
while another step in CV)