Store Operation
To save the load’s settings to a register (internal memory), press the and
(Save) keys. The LEDs will blink together.
Press the or or or keys to save the settings to non
volatile memory. Pressing A will store to A location. B to B location. C to C
location. D to D location. Hence, users can save up to four settings in each
group. Note you may do this for any of the 40 different groups assigned in the
previous paragraph (Choose Storage Group). Each settings can store two
parameters: The mode of operation (i.e. CC, CV, CR) and the values set for that
mode.
ShiftShift
AA
AA
BB
View
C
Mode
D
Mode
D
A
B
C
D
Short
When the + keys are pressed, the load simulates a short circuit. In
CC and CR modes, the maximum short-circuit current value is 1.2 times the
current range. In CV mode, the short-circuit current is the same as setting the
CV operation to 0.1 V. Note that the short-circuit current in CV mode will be
less than the short-circuit current in CC or CR modes.
ShiftShift
On/OffOn/Off
Recall Operation
Press the and (Recall) keys. The Recall LED will turn on, which
means you may recall settings from the currently-chosen group. Press the
or or or keys to recall the desired setting. When the Recall
LED is lit, the other keys are disabled. To exit the recall state, press the
and (Recall) keys again.
ShiftShift
BB
ShiftShift
BB
AA
BB
View
C
Mode
D
Mode
D
The and keys switch between the 3 A and 30 A ranges. The
30A LED is lit when the load is on the 30 A range. The current resolution is
10 mA for the 30 A range and 1 mA for the 3 A range.
Current range switch (3/30 A)
ShiftShift
Mode
D
Mode
D
Checking the set value
The electronic load usually displays the actual voltage and current values at
the load’s terminals. To check the set value, press the or keys. The
set value display will blink, showing you the currently-set value. You can use
the knob to change it. It will stop blinking after 3 seconds and revert to the
previous display. You can also turn the knob to see the set value.
AA
BB
Verify Firmware
To verify the firmware version of your DC Electronic load, do the following:
- First, power off the unit. Then hold down the button.
- Turn the power switch ON while holding down the button.
- The top display will show “VEr” and bottom display will show a number,
which is the firmware version of your load.