“Step Delay” This is the delay time to hold each power step in the test. This determines how fast or
slow to run through the test.
“End Power” This is the ending power value before OPP test ends. The test will only run through in steps from
“Start Power” to the “End Power” if input voltage is greater than the “OPP Voltage”.
“OPP Voltage” This value is a limit that is used to test for OPP conditions. In the event of an over power
condition, when the input voltage is greater than this value, the test will run through from “Start
Power” to “End Power” in steps. The input power (the over power) value will be checked during
this process. If power is within the “Max Trip Power” and “Min Trip Power” boundaries, the load
will indicate “Pass” at the end of the test. If outside of these boundaries, the load will indicate
“Fault”.
“Max Trip Power” The maximum acceptable power limit of the over power value.
“Min Trip Power” The minimum acceptable power limit of the over power value.
Operations to clear the OVP state
Disconnect the instrument under test. Press any key on the front panel, the OVP on the VFD will disappear, then the
DC load exits OVP protection state.
5 Over Current Protection (OCP)
The DC load includes both hardware and software over current protection features.
Hardware OCP: maximum input current of the DC load will be limited at about 110% of the current range, once the
hardware OCP is activated, the status register’s OC bit will be set; when the hardware OCP is removed, the status
register’s OC bit will be reset. Hardware over current protection won’t change the DC load’s On/O state.
Software OCP: users can set the DC load’s software OCP value, steps:
[Shift] + [9] > Protect > Alimit set ON, Apoint set OCP current value, Adelay set delay time before alarm. When the
software OCP function is active, the DC load will automatically turn o, VFD displays OCP.
Operations to clear the OCP state
Disconnect the instrument under test. Press any key on the front panel, the OCP displayed on the VFD will disappear,
the DC load exits OCP protection state.
6 Over Power Protection (OPP)
The DC load includes both hardware and software OPP features.
Hardware OPP: the DC load allows user to set a power protection limit in hardware which will limit the power in the
range you set when the OPP occur. The hardware OPP protection will not change the ON/OFF state of the the DC
load.
Software OPP: users can set the DC load’s software OPP value, steps: [Shift] +[9] > Protect > P-LIMIT > P-LIM
POINT set OPP power value, P-LIM DELAY set alarm delay. If the the DC load’s power value reach OPP limit and
after the delay time, the DC load will automatically turned o, VFD will display OPP.
Operations to Clear the OPP State
Disconnect the instrument under test. Press any key on the front panel, the OPP displayed on the VFD will disappear,
the DC load exits OPP protection state.
7 Over Temperature Protection (OTP)
If internal temperature exceeds safety limits(85C ;185F), the Over temperature circuitry will be activated. The DC Load
will turn o the input, the buzzer will sound, and the display will show OTP.
Operations to Clear the OTP State
When the DC load temperature drops to the protecting point, press any key on the front panel, the OTP) displayed on
the front panel will disappear, the DC load exits OTP protection state.