4. It will then prompt to enter a value for “Delay”. This is the amount of time to enable the input from when the is
pressed (or when input is turned ON remotely). Once this time is past, input will be disabled (OFF). Use the numeric
keypad or rotary knob to enter the value. Press to conrm the change. The valid range is 0 – 60000 seconds.
5. Now, press twice to exit the menu. The instrument will show the delay timer as follows:
8.5.22 Measurement Congurations
Voltage Auto Range
By default, the load has auto range enabled for voltage measurements and operations. To enable or disable, follow the
steps below:
1. Go to “CONFIG” menu and select “Measure” and press .
2. Select “V-Range” and press .
3. To enable voltage auto range, select “On” and press . To disable, select “O” and press to conrm the
change.
Measuring Rise and Fall Time
The instrument can measure the rise or fall time from a specied start and stop voltage level of the measured input.
This feature requires the display timer to be enabled rst. To enable timer, please follow the instructions in the “Display
Input On Timer” section of “SYSTEM Menu”.
To setup this measurement, follow the steps below:
1. Go to “CONFIG” menu and select “Measure”. Then press .
2. Select “TimeV1” and press . The load will prompt to enter a value. Use the numeric keypad or rotary knob to
set a value and press to conrm. This is for the start voltage level. When the measured input voltage reaches
this level, the timer will start.
3. Now, select “TimeV2” from the “Measure” menu and press . The load will prompt to enter a value. Use the
numeric keypad or rotary knob to set a value and press to conrm. This is for the stop voltage level. When
the measured input voltage reaches this level, the timer will stop. The time on display will show the time dierence
between the measured start and stop voltage level.
Measurement Averaging Filter
The measurement averaging lter is adjusted. Increasing the averaging will provide more accurate readings, but slower
measurement update rate. Decreasing the averaging will provide faster measurement update rate, but less accurate
readings.
To congure, follow the steps below:
1. Go to “CONFIG” menu and select “Measure”. Then press .
2. Select “Filter” and press . The load will prompt to enter a value for “Average Count”.
3. Use the numeric keypad or rotary knob to adjust this value, then press to save the change. The valid range is
from 2 – 16 (2^2 – 2^16). The default is 14 (2^14).