User Manual
39
Storage Group Memory locations
0 1 – 10
1 11 – 20
2 21 – 30
3 31 – 40
4 41 – 50
5 51 – 60
6 61 – 70
7 71 – 80
8 81 – 90
9 91 - 100
When recalling settings, each of the numeric keypad numbers corresponds to the memory locations based on the storage
group selected according to the table above. For storage group 0, recalling memory location 1 is done by pressing ;
location 2 is done by pressing , and so on. Memory location 10 is recalled by pressing . For storage group 1, recalling
memory location 11 by pressing , location 12 by pressing , and so on.
Example:
Settings are saved to memory location 60. To recall those settings, set storage group to 5 from the menu, then press
recall and the number .
8.5.5 Select Storage Group
1. From the “SYSTEM” menu, browse and select “Memory” and press . The following screen will appear.
1. Use the current adjust knob or the numeric keypad to enter the storage group. Select between “0 – 9”. Press to
save selection.
2. To exit the menu at any time, press twice.
NOTE: The storage group setting also aects the automatic test function of the load. Refer to “Automatic Test
Function” for more details.
8.5.6 Save Settings
1. Set up all the instrument settings that you want to save.
2. Then, press ) and . The display will show the following:
3. Use the current adjust knob or the numeric keypad to enter the memory location in which to store current instrument
settings. Select between “0 – 100”. Press to save to the selection location.
NOTE: The “0” memory location is reserved for storing instrument settings last congured before power-o and is used
only for power-on state conguration only.
8.5.7 Recall Settings