User Manual

36
System Settings
8.1 Run Recall Edit
1. Select “Edit” and press . Enter a value using the numeric keypad or rotary knob for “Voltage On Level”. The value
must be within the maximum input limits of the load. Press to continue.
2. Enter a value for “Voltage On Delay”. This is set between 0.00s and 99.99s. Then press to continue.
1. Follow the same steps for “Current Range”. The value must be within the maximum input limits of the load. Press
to continue.
2. Set the “Start Power”. The value must be within the maximum input limits of the load. Press to continue.
3. Follow the same steps for “Step Power” and “End Power”. Press after each settings to continue.
4. Enter a value for “OPP Voltage” and press to continue.
5. Enter a value for “Max Trip Power”, and then “Min Trip Power”. Press after each settings to continue.
1. The load will prompt to “Save OPP File”. Select a number between 1 and 5. Then press to save all settings to the
selected location.
Note: At any time when conguring any parameters for the test, you can press the ) key to select the previous parameter
to edit.
8.2 To recall the settings
1. Press to enter the “OPP Test” menu.
2. Select “Recall” and press .
3. Use the keypad or knob to select the saved location number and press . All settings will be recalled when selecting
“Edit” from the “OPP Test” menu.
8.3 To run the OPP test
1. Recall the settings from memory by following the steps above.
2. From the “OPP Test” menu, select “Run” and press . The display will show the following:
OFF CC
v-0.000a 10.000V 0.000A
0.00W 0.00W Stop
3. Press ) to start the test. When the test is running, “Run” will be indicated in place of “Stop”. When the test ends,
either “Pass” or “Fault” will appear next to “Stop”.
4. To stop the test at any time, press ).
8.4 Key Lock
The front panel keys are locked to prevent unwanted changes to output settings and instrument congurations. Follow
the steps below to enable/disable key lock.