User Manual

24
Figure 6.1 List Mode Current Waveform
6.4.1 Congure List
Lists are run exclusively from internal memory only. As in other menus, an active eld is denoted by an underline cursor
or ashing. To enter a list into memory, complete the following:
1. Press + .
2. Set the rst eld in the menu to “OFF”. If the eld is set to “ON”, press to change the value to “OFF”.
3. Move the selected eld via the buttons to the “EDIT” (shown as EDI) eld and press to start editing a
List. If nothing happens, ensure the rst eld is set to “OFF”.
4. Set the current range.
5. Set the number of steps in the list. This is a value between 2 and 84.
6. Set the rst step’s current, slew rate, and step time.
7. The next step in the list will prompt automatically. Enter the values for each step in the same way.
8. When each step is entered, the system prompts for the number of times to run the sequence. A 1 in this eld means
the sequence will run only once.
9. Enter the list le location to store the newly created list.
The list is now saved and may be recalled.
6.4.2 Run List
1. Select the list to run:
1. From the “List” menu, select “CALL” and press .
The load prompts for the list to recall (1 through 7).
2. Enter the list number and press to recall it.
2. Enable List mode by selecting the “O” menu item and press .
The “OFF” changes to “ON” and the “Trig” indicator lights.