Battery Capacity Analyzer User Manual
37
Press OPTION to TRG
Arrow Keys to Set
I Min = 00323 mA
I Med = 07869 mA
6.4 Load Calibration
ThisisasimilarprocesstoCurrentcalibrationwithonly3calibrationpoints.Thebuttonisusedtotriggerthe603B
placingaloadonthepowersupply.Monitorthemaximumcurrentusedasdisplayedonthepowersupplycurrentmeter.
Writethemaximumvaluedown.Enterthis4digitvalueintherow.Repeatuntilallthreecalibrationpointsareadjusted.
Notethatthevalueinthe603BisinmA.
Fromthe3partcalibrationscreenmovethepointtotherow“LoadCalibration”.
25. PressthebuttontoselectLoadcalibration
26. Whenready,pressthebuttontotriggerthe603B.
27. UsethebuttontomovethecursorintotheLoadeldandenterthemaxcurrentdisplayedonthepowersupply.
28. RepeatthesestepsuntilallMin,Med,Maxcurrentvalueshavebeencalibrated.Notethisisspreadacross
2 separate screens.
29. Completeafantestandverifythe603Bfanrunssmoothly.
30. Createacalibrationlabelandapplytothe603B.
The Calibration procedure is complete.
I Med 4 = 01462 mA
I Max 4 = 02896 mA
Current Calibration 4 of 4 screens
Voltage Calibration
Current Calibration
Load Calibration
Load Calibration 1 of 2
I Max = 11350 mA
Load Calibration 2 of 2