16. Pressthebuttontoselectcurrentcalibration.
17. Setthevoltageonthepowersupplyto12.000volts.
18. Setthecurrentonthepowersupplyto12.000amps.
19. SetthepowersupplytoMetermodesothecurrentcanbemonitored.
20. Verifythepointerislocatedontherstrow,“I_Min...”
Thenextfewstepsareusedtocalibratethe603Bcurrent12calibrationpointsareusedover3screenviews.
21. Thebuttonisusedtotriggerthe603Bplacingaloadonthepowersupply.Monitorthemaximumcurrentused
asdisplayedonthepowersupplycurrentmeter.Makenoteofthemaximumvalue.Enterthis4digitvalueinthe
row.Repeatuntilall12calibrationpointsareadjusted.Notethatthevalueinthe603BdisplayisinmAandmost
power supplies display in amps so a conversion will be required.
22. Whenready,pressthebuttontotriggerthe603B.
23. Usethebuttontomovethecursorintothecurrenteldandenterthemaxcurrentdisplayedonthepower
supply.
24. Repeatthesestepsuntilall12currentvalueshavebeencalibrated.Notethisisspreadacross4separatescreens.
Voltage Calibration
Current Calibration
Load Calibration
Current Calibration 1 of 4 screens
I Max 1 = 02904 mA
I Min 2 = 00092 mA
I Med 2 = 01474 mA
I Max 2 = 02916 mA
Current Calibration 2 of 4 screens
I Min 3 = 00092 mA
I Med 3 = 01462 mA
I Max 3 = 02894 mA
I Min 4 = 00092 mA
CurrentCalibration3of4screens
Press OPTION to TRG
Arrow Keys to Set
I Min 1 = 00092 mA
I Med 1 = 01466 mA
6.3 Current Calibration
15. Fromthe3partcalibrationscreenmovethepointtotherow“CurrentCalibration”.