User Manual
Battery Capacity Analyzer User Manual
35
9. Verifythepointerislocatedattherstrow“V_Min_1…”
10. PressthebuttontosettheADCoset.Thenormalfunctionisthe4digitvalueaftertheequalssignwillchange.
It is also normal if this value does not change as correction was not required.
11. FromtablexxxndthenextvoltageforV_Med_1of8.00V.
12. Movethepointertothenextrowdown“V_Med_1…”andpressthebutton.
13. ContinuethesameprocessfortheRemainingvoltagesettingsuntilall6voltshavebeencalibrated.
14. Continue to the Current calibration section
Press ENTER to Set
V Min 1 = 1114 ADC
V Med 1 = 1627 ADC
V Max 1 = 3607 ADC
Voltagescreen1of2
603B Row
Label
Voltage
Setting
VMin1 5.40V
VMed1 8.00V
VMax1 18.00V
V Min 2 = 0048 ADC
V Med 2 = 2135 ADC
V Max 2 = 3596 ADC
Voltagescreen2of2
603B Row
Label
Voltage
Setting
VMin2 15.00V
VMed2 25.00V
VMax2 32.00V
6.2 Voltage Calibration
Thissectiondescribeshowtocalibratethevoltage.ThelabeliseasytoconfuseasV_Min_1issimilartoV_Med_1
carefullymatchthevaluesinthetablestothelabelonthe603Bateachstep.
FromtableXXXndthevoltageforV_Min_1of5.40Vandsetthepowersupplytothisvoltage.