User Manual

67
DC Current: ± (% of reading + % of range), 23 °C ± 5 °C
Rate
Range
Resolution
Full Scale
Reading
Accuracy
(1 year)
Burden Voltage
(1)
& Shunt Resistor
Slow
2.0000 mA
20.000 mA
200.00 mA
2.0000 A
20.000 A
0.1 µA
1 µA
10 µA
100 µA
1 mA
2.1000
21.000
210.00
2.1000
21.000
(3)
0.08%+0.025%
(2)
0.08%+0.02%
(2)
0.08%+0.02%
0.3%+0.025%
0.3%+0.025%
<0.3 V / 100 Ω
<0.04 V / 1 Ω
<0.3 V / 1 Ω
<0.05 V / 10 mΩ
<0.6 V / 10 mΩ
(1)
Typical voltage across the input terminals at full scale reading.
(2)
Use REL function
(3)
In 20 A range, >10-20 ADC is readable for 20 seconds maximum
Maximum Input and Overload Current Protection: 1 A/ 250 V fuse.
AC Current (True RMS, AC Coupling)
ACCURACY: ± (% of reading + % of range), 23 °C ± 5 °C
Rate
Range
Accuracy(1 year)
(1)
23 °C ± 5 °C
20-50 Hz
50-2 kHz
2-20 kHz
Slow
2.0000 mA
20.000 mA
200.00 m A
2.0000 A
20.000 A
1.5%+0.5%
1.5%+0.5%
1.5%+0.5%
2.0%+0.5%
2.0%+0.5%
0.5%+0.3%
0.5%+0.3%
0.5%+0.3%
0.5%+0.5%
0.5%+0.5%
2%+0.55%
2%+0.38%
2%+0.5%
------
------
(1)
Specifications are for sine wave inputs >5% of range.
Measurement method: True RMS measurement
Maximum Crest factor: 3.0 at full scale
Maximum Input and Overload Current Protection: 1 A/ 250 V fuse.
Rate
Range
Resolution
Full Scale
Reading
Burden Voltage
(1)
/shunt Resistor
Slow
2.0000 mA
20.000 mA
200.00 m A
2.0000 A
20.000 A
0.1 µA
10 µA
100 µA
1 mA
10 mA
2.1000
21.000
210.00
2.1000
21.000
(2)
<0.3 V / 100 Ω
<0.04 V / 1 Ω
<0.3 V / 1 Ω
<0.05 V / 10 mΩ
<0.6 V / 10 mΩ
(1)
Typical voltage across the input terminals at full scale reading.
(2)
In 20 A range,>10-20 A AC is readable for 20 seconds maximum