CHAPTER 8. INTEGRAL OPERATION FUNCTION 41
8.5 Integration
When the integral measurement function is enabled, you can keep the current integral information and carry out
the following operations: exit, start and stop. Specic steps are as follows:
Operation steps
1. Press “Integ” to enter the integral display interface.
2. In the integral measurement display interface Press the soft key corresponding to the “START” parameter
to start or recover integral operation. The Start status and integral operation value are displayed in the
interface.
3. When you need to keep the current integral status and integral result, you can press the soft key
corresponding to the “RUN HOLD” parameter. Then the current status of the integral result display and
communication output will be kept no matter whether integral operation is running.
4. When you need to continue to implement integral operation, double-press the soft key corresponding to
the “RUN HOLD” parameter to exit the HOLD status. The power meter will display the integral result
after exiting the HOLD status. The schematic diagram of display of the HOLD/RUN integral result is
shown below.
5. When you need to suspend integral operation, press the soft key corresponding to the “Stop” parameter to
suspend integral operation.
6. Press the soft key corresponding to the “RESET” parameter to reset the integral value and integral time.
The relationship diagram of start, suspension and resetting of the integral operation status is shown below.
Auto Clear when ON is selected.
Auto Clear when OFF is selected.