User Manual

Chapter 4
Measurement Setup
Central to setting the proper range of the instrument is the Crest Factor. Crest factor is the ratio of the peak
value of a waveform to the RMS value of the waveform. For example, a perfect sine wave crest factor RMS
value is 0.707*Peak. The crest factor is the inverse of the normalized RMS value, 0.707
1
= 1.414. In many
cases, like the current waveform of a AC-DC power supply, the crest factor is a larger value. For accurate
measurement, the signal measured should stay within the measurement range of the instrument. For example, a
100V RMS signal with a crest factor of 5 has peaks of 500V. If the range selected is smaller than 500V than
inaccuracy will occur as the signal will be clipped.
4.0.1 Crest factor
The Crest Factor, the ratio of the peak to the RMS value, has two settings, 6 or 3. This setting applies to both
current and voltage simultaneously, and modies the available ranges listed in Section 4.0.2. Essentially, by
knowing the RMS signal level and the peak value, the setting of this parameter is determined. For an 117V
RMS AC line signal with little distortion and a purely real load (power factor = 1), the peak value is
1.414 117V = 165V . 1.414 is the approximate Crest Factor of a Sine wave. So, the appropriate setting will be
3 (CF3).
CrestF actor(CF ) =
P eak
RM S
As this meter also measures the DC level of the signal, this must also be accounted for when choosing the Crest
Factor. If the the signal will exceed the range of the unit (crest factor * range) then the measurement value will
be clipped and the value inaccurate. Conversely, by choosing a Crest Factor and Range combination that
accommodates the peak value of the signal, a larger crest factor than set may be measured. The trade-o is
then the accuracy, and is listed in the specications, Chapter 10.
Example (Range=150 V, CF = 6): The maximum input signal and crest factor are as follows:
V
rms
= 100 Range = 150V V
maxrange
= 900V
So, a signal with a peak that is 9 times that of the RMS value (Crest Factor 9) is measurable using these
settings.
4.0.2 Set Measurement Range
The appropriate measurement range (voltage and current range) must be set for accurate measurement.
18