CHAPTER 2. MENU 13
Figure 2.12: Ratio Set Menu
2.2.3.1 Filter
Frequency lter it is inserted in the frequency measurement circuit and may aect frequency measurement. It
can be used for ltering high-frequency components of interference to make the measured frequency
parameter more accurate. When the frequency lter is switched on, the voltage or current of no more
than 200Hz can be measured. The cuto frequency is 500HZ.
Line lter it is inserted in the voltage and current measurement circuit and has direct inuence on measurement
of the voltage, current and power. When the line lter is switched on, noise and high-frequency
components from the inverter or distortion waveform can be ltered. The cuto frequency is 500HZ.
2.2.3.2 Crest factor
The crest factor is the ratio of the waveform peak to the eective value. The Measurement Conditions crest
factor of 5335B is specied as the times of the crest value which can be input under the rated input conditions.
The crest factor CF3 or CF6 can be selected in the interface “Menu > SETUP > OTHER SET”. See Chapter 4
for more details.
2.2.4 Inrush Measurement Setup
Feature not currently implemented.