User Manual
Sine Waveform 18
2.5.2 Harmonic Order
After entering the harmonic setting menu, press Order, the use the numeric keyboard or knob to input the desired value.
• The range is limited by the maximum output frequency of the instrument and current fundamental waveform frequency.
• Range: 2 to maximum output frequency of the instrument ÷ current fundamental waveform frequency The maximum
is 10.
2.5.3 Harmonic Amplitude
After entering the harmonic setting menu, press Harmonic Ampl to set the harmonic amplitude of each order.
1. Press Order to select the sequence number of the harmonic to be set.
2. Press Harmonic Ampl to set the amplitude of the harmonic selected. Use the arrow keys and knob to change the
value. Or use the numeric keyboard to input the amplitude value and then select the desired unit from the pop-up
menu. The units available are Vpp, mVpp and dBc.
2.5.4 Harmonic Phase
After entering the harmonic setting menu, press Harmonic Phase to set the harmonic phase of each order.
1. Press Order to select the sequence number of the harmonic to be set.
2. Press Harmonic Phase to set the phase of the harmonic selected. Use the arrow keys and knob to change the value.
Or use the numeric keyboard to input the phase value and then select the unit °.