4050B Series Function/Arbitrary Waveform Generator USER MANUAL i
Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument. Before applying power to this instrument: Read and understand the safety and operational information in this manual. Apply all the listed safety precautions. Verify that the voltage selector at the line power cord input is set to the correct line voltage.
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to the primary power entering a building or other outdoor wiring. Do not use this instrument in an electrical environment with a higher category rating than what is specified in this manual for this instrument. You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the instrument's category rating to maintain the instrument's category rating.
Unless otherwise stated, a ground connection on the instrument's front or rear panel is for a reference of potential only and is not to be used as a safety ground. Do not operate in an explosive or flammable atmosphere Do not operate the instrument in the presence of flammable gases or vapors, fumes, or finelydivided particulates. The instrument is designed to be used in office-type indoor environments.
Measurements made by this instrument may be outside specifications if the instrument is used in non-office-type environments. Such environments may include rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise, strong electric fields, or strong magnetic fields.
Do not touch live circuits Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must be made by qualified service-trained maintenance personnel who are aware of the hazards involved when the instrument's covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may exist when the covers are removed.
Cooling fans This instrument contains one or more cooling fans. For continued safe operation of the instrument, the air inlet and exhaust openings for these fans must not be blocked nor must accumulated dust or other debris be allowed to reduce air flow. Maintain at least 25 mm clearance around the sides of the instrument that contain air inlet and exhaust ports. If mounted in a rack, position power devices in the rack above the instrument to minimize instrument heating while rack mounted.
Compliance Statements Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European countries with separate collection systems) This product is subject to Directive 2002/96/EC of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste.
CE Declaration of Conformity This instrument meets the requirements of 2006/95/EC Low Voltage Directive and 2004/108/EC Electromagnetic Compatibility Directive with the following standards.
Safety Symbols Refer to the user manual for warning information to avoid hazard or personal injury and prevent damage to instrument. Electric Shock hazard Alternating current (AC) Chassis (earth ground) symbol. Ground terminal On (Power). This is the In position of the power switch when instrument is ON. Off (Power). This is the Out position of the power switch when instrument is OFF. Power Switch (On/Off). This is the power switch located in front of the instrument.
Table of Contents SAFETY SUMMARY .................................................................................................................................................. I COMPLIANCE STATEMENTS ............................................................................................................................................ VII SAFETY SYMBOLS ....................................................................................................................................................
Selecting Modulation Source .................................................................................................................................... 43 To Set Modulation Depth .......................................................................................................................................... 44 To Set Modulation Frequency ...................................................................................................................................
4.4 4.5 OUTPUT ...........................................................................................................................................................77 Load ...............................................................................................................................................................78 Polarity ..........................................................................................................................................................
Example 8: Generate a Burst Waveform .....................................................................................................110 Example 9: Generate an AM Modulation Waveform ..................................................................................112 Example 10: Generate an FM Modulation Waveform ................................................................................113 Example 11: Generate a PM Modulated Waveform ..............................................................
1 General Information 1.1 Product Overview BK Precision’s 4050B series are dual-channel function/arbitrary waveform generators, capable of generating sine and square waves of up to 10 MHz, 30 MHz or 60 MHz (depending on model). They have an informative easy-to-read color display, user-friendly controls and a numeric keypad which allows users to easily configure waveform properties.
1 x 4050B series function/arbitrary waveform generator 1 x Getting started manual (printed) 1 x AC power cord 1 x USB type A to Type B cable 1 x Certificate of calibration Verify that all items above are included in the shipping container. If anything is missing, please contact B&K Precision. An optional USB to GPIB adapter model AK40G is also available.
1.3 Front Panel Overview 10 9 8 11 7 6 1 2 3 4 5 Front Panel Description 1 Power On/Off switch 2 Menu softkeys 3 Menu buttons 4 Channel 1 output terminal 5 Channel 2 output terminal 6 Channel 1 and 2 Output On/Off buttons 7 Arrow keys 8 Rotary dial knob 9 Numeric keypad 10 TFT LCD color display USB host port/*USB-to-GPIB adapter interface 11 Accepts USB flash drive to save/recall instrument settings and waveforms.
1.
1.
2 Getting Started Before connecting and powering up the instrument, please review and go through the instructions in this chapter. 2.
1. Use good quality 50 Ω coaxial cable and connectors. 2. Make all connections tight and as short as possible. 3. Use good quality attenuators, if it is necessary to reduce waveform amplitudes applied to sensitive circuits. 4. Use termination or impedance-matching devices to avoid reflections. 5. Ensure that attenuators and terminations have adequate power handling capabilities. If there is a DC voltage across the output load, use a coupling capacitor in series with the load.
Figure 1 - Preliminary Self-Test Figure 2 - Self-Test Options Note: Refer to TEST/CAL section for more information about the Self-Test function. Check Model and Firmware Version The model and firmware version can be verified from within the menu system. Press Utility, press System select and press Page 1/2 from the menu to enter the second menu page, and select System Info option. The software/firmware version, hardware version, model, and serial number will be displayed.
Output Check Follow the steps below to do a quick check of the settings and waveform output. 1. Turn on the instrument and set the instrument to default settings. To set to default, press Utility → System → Set To Default to set the system to the default setting.
Start Phase 0° Cycles 1 Cycle Trig Out Off Delay 521 ns Trigger Default Source Internal Table 1 - Default Settings 1. Connect the BNC output of CH1 (yellow) into an oscilloscope. 2. Press the Output button on top of CH1 output BNC to turn on the output and observe a sine wave with the parameters above. 3. Press the Parameter button. 4. Press Freq or Period option in the menu and use the rotary knob or the numeric keypad to change frequency. Observe the changes on the oscilloscope display. 5.
3 Front Panel Operation 3.1 Menu Options All settings and parameters can be configured from the menu system of the instrument. The channel specific menu options are the same for both channel 1 and channel 2. Use the Ch1/Ch2 button keys to toggle the channel selection. The selected option will be highlighted in blue. Some settings are common for most waveforms (i.e. frequency, amplitude, offset, etc.) and some are specific to each type of waveform (i.e.
Offset/LowLevel Configures the DC offset or the low level of the waveform. Phase Configures the phase relative to the other channel. DutyCycle Configures the duty cycle of the waveform. Frequency/Period Configures the frequency or period of the waveform. Amplitude/HighLevel Configures the amplitude or the high level of the waveform. Offset/LowLevel Configures the DC offset or the low level of the waveform. Phase Configures the phase relative to the other channel.
Source Selects modulating source. AM Depth Set the modulation depth. Shape Configures the modulating waveform shape. AM Freq Set the modulating waveform frequency. Frequency range: 1 mHz to 20 kHz (internal source only). Sweep Time Configures the sweep time for sweep output. StartFreq/CenterFReq Configures the sweep start frequency or the center frequency. StopFreq/FreqSpan Configures the sweep stop frequency or the frequency span of the sweep output.
Output Setup Set the output parameters of CH1 and CH2. CH Copy Coupling Set the track, channel coupling or channel copy function. Interface Select and set parameters of remote communication. Clock Choose the system clock source: internal or external. Mode Choose Phase-locked or independent mode. OverVoltage Protection Turn on/off the overvoltage protection function. Table 3 - Menu System Organization CHCopy Copy channel settings between channels. 3.
Figure 4 - Channel 2 Selected 3.3 Configure Waveform Output Configure Waveform Shape The instrument can generate many standard as well as arbitrary waveforms. There is a dedicated waveform key on the front panel that will allow the user to select between different waveform shapes to output, as listed in Table 4.
Press Waveforms to select a waveform. Menu options relevant to the selected waveform shape will display at the bottom of the screen. The screenshots below illustrates the menu options for each of the waveform types.
Table 5 - Waveforms Configure Frequency This section does not apply to noise or DC waveforms. Press the Frequency/Period softkey to toggle between the Frequency and Period settings. The option selected will be highlighted in blue. The current value for the waveform’s frequency or period is now highlighted on the Waveform Parameters display section of the screen. The adjustable frequency range is different for each model and for each type of waveforms.
4. Use the menu function keys to select the unit. Available units are: MHz, kHz, Hz, mHz, and uHz. Note: When using the numeric keyboard to enter the value, the left arrow key can be used to move the cursor backward and delete the previous digit. Configure Amplitude This section does not apply for noise or DC waveform.
User-Defined High and Low Level The user has the option to adjust the high and low level of the waveform. To do this, toggle the menu function key from Amplitude to HighLevel and the LowLevel option will be available as well. Figure 7 - Amplitude: High and Low Level Note: DC offset settings will be automatically adjusted if the high and low levels reflect a DC offset. Configure Offset This section does not apply to noise waveform.
Figure 8 - Offset 6. Use the menu function keys to select the unit. Available units are: Vdc and mVdc. Configure Phase This section does not apply for pulse, noise, or DC waveforms. The Phase setting is useful for adjusting the phase relationship between channels 1 and 2, between a channel and its sync signal, and synchronizing multiple instruments. This setting is different from the Start Phase in Burst mode. Follow the steps below to configure the phase of the output: 1. 2. 3. 4. 5.
Figure 9 - Phase 6. Use the menu function keys to select the degree (°) unit. Note: When the IndependentMode is enabled, the phase parameter cannot be modified (i.e., the phase can only be set when there is a reference clock waveform provided on the back panel). Configure Duty Cycle: Square Waveform Figure 10 - Duty Cycle The duty cycle setting range is limited by the “Frequency/Period” setting. The default value is 50%. Follow the steps below to configure the phase of the output: 1.
Figure 11 - Duty Cycle of a Square Waveform 4. Use the menu function keys to select the degree (°) unit. Configure Width and Duty Cycle: Pulse Waveform The pulse width and duty cycle parameters are related to each other and both control the length of the “On Time” of a pulse. Users have the option to specify the pulse width in units of seconds or the duty cycle as a percentage. Pulse duty cycle and pulse width are dependent— changing one of the parameters will automatically change the other.
Follow the steps below to configure the pulse width or duty cycle. 1. Press the Waveforms button. 2. Press the Pulse softkey. 3. Select PulWidth softkey for pulse width adjustment or DutyCycle for duty cycle adjustment. 4. The cursor position will now highlight the first digit of the width or duty parameter display. Figure 13 - Pulse Width or Duty Cycle 5. Use the rotary knob or the numeric keypad to change the width or the duty.
3. Toggle the Rise/Fall softkey to set the rise or fall edge. The Rise/Fall softkey lets the user toggle between the Rise and Fall Edge settings. The selected setting will be highlighted in blue. Use the numeric keyboard to input the parameter value directly and press the corresponding key to select the parameter unit. Alternatively, use the arrow keys to select the digit and use the knob to change its value.
Figure 16 - Pulse Delay 4. Use the rotary knob or the numeric keypad to change the pulse delay. After entering the numeric value, use the menu function keys to select s, ms, us, or ns. Configure Symmetry: Ramp Waveform A ramp waveform with 50% of symmetry is a triangular waveform. A sawtooth waveform is a ramp waveform with either 0 or 100 percent. Figure 17 - Symmetry Follow the steps below to configure the symmetry of a ramp/triangle waveform. 1. Press the Waveforms button. 2.
Figure 18 - Ramp Symmetry 4. Use the rotary knob or the numeric keypad to change the symmetry. 5. Once the value has been entered, press the “%” softkey from the menu to set the value. Configure Standard Deviation and Mean: Noise Waveform There are two parameters that can be adjusted of the noise waveform: Standard deviation and mean. Noise is a non-periodic random signal which has no frequency or period. Follow the steps below to configure these parameters. 1. Press the Waveforms button. 2.
Figure 19 - Noise Standard Deviation Figure 20 - Noise Mean 5. Use the rotary knob or the numeric keypad to change the two parameters. Both parameters can be specified in V or mV units. Configure Offset: DC Waveform The instrument can output a DC waveform output at a range of voltage levels (-12 Vdc – 12 Vdc). There is only one parameter, DC Offset, to configure a DC output waveform. Follow the steps below to configure the parameter. 1. Press the Waveforms button.
2. Select and press the DC waveform softkey from the menu. 3. The cursor position will now be highlighted the first digit of the DC offset parameter display. Figure 21 - DC Offset 4. Use the rotary knob or the numeric keypad to change the two parameters. Both parameters can be specified in Vdc or mVdc units. Configure Arbitrary Waveforms There are two ways to generate arbitrary waveforms.
Waveform Function Waveform Function StairUp Stair-up waveform Npulse Negative pulse StairDn Stair-down waveform UpRamp UpRamp waveform StairUD Stair-up and down waveform DnRamp DnRamp waveform Trapezia Trapezia waveform SineTra Sine-Tra waveform Ppulse Positive pulse SineVer Sine-Ver waveform Arbitrary Waveform: Math Waveform Function Waveform Function ExpFall ExpFall function Legend 5-times Legend polynomial ExpRise ExpRise function Versiera Versiera LogFall LogFall fun
Arbitrary Waveform: Engine Waveform Function Waveform Function Cardiac Cardiac signal SCR SCR firing profile Quake Analog quake waveform TV TV signal Chirp Chirp signal Voice Voice signal TwoTone TwoTone signal Surge Surge signal SNR SNR signal Radar Analog radar signal AmpALT Gain oscillation curve Ripple Ripple wave of battery AttALT Attenuation oscillation curve Gamma Gamma signal RoundHalf RoundHalf Waveform StepResp Step-response signal RoundsPM RoundsPM Waveform B
Kaiser Kaiser window BohmanWin BohmanWin window Blackman Blackman window ChebWin ChebWin window GaussiWin GaussiWin window FlattopWin Flat top weighted window Triangle Triangle window (Fejer window) ParzenWin ParzenWin window BlackmanH BlackmanH window TaylorWin TaylorWin window Bartlett-Hann Bartlett-Hann window TukeyWin TukeyWin (tapered cosine) window Arbitrary Waveform: Trigonometric (Trigo) Waveform Function Waveform Function Tan Tangent Csch Hyperbolic cosecant Cot Cot
SquareDuty02 Square waveform with 2% duty SquareDuty38 Square waveform with 38% duty SquareDuty04 Square waveform with 4% duty SquareDuty40 Square waveform with 40% duty SquareDuty06 Square waveform with 6% duty SquareDuty42 Square waveform with 42% duty SquareDuty08 Square waveform with 8% duty SquareDuty44 Square waveform with 44% duty SquareDuty10 Square waveform with 10% duty SquareDuty46 Square waveform with 46% duty SquareDuty12 Square waveform with 12% duty SquareDuty48 Square
Arbitrary Waveform: Square 2 Waveform Function Waveform Function SquareDuty70 Square waveform with 70% duty SquareDuty86 Square waveform with 86% duty SquareDuty72 Square waveform with 72% duty SquareDuty88 Square waveform with 88% duty SquareDuty74 Square waveform with 74% duty SquareDuty90 Square waveform with 90% duty SquareDuty76 Square waveform with 76% duty SquareDuty92 Square waveform with 92% duty SquareDuty78 Square waveform with 78% duty SquareDuty94 Square waveform with 94%
ECG3 Electrocardiogram 3 ECG15 Electrocardiogram 15 ECG4 Electrocardiogram 4 LFPulse Waveform of the low frequency pulse electrotherapy ECG5 Electrocardiogram 5 Tens1 Waveform 1 of the nerve stimulation electrotherapy ECG6 Electrocardiogram 6 Tens2 Waveform 2 of the nerve stimulation electrotherapy ECG7 Electrocardiogram 7 Tens3 Waveform 3 of the nerve stimulation electrotherapy Arbitrary Waveform: Mod Waveform Function Waveform Function AM Sectional sine AM signal PM Sectional si
Follow the steps below to browse and select a predefined arbitrary waveform. 1. Press the Waveforms button. 2. Go to the second page of waveform options by pressing the 1/2 softkey. 3. Select and press the Arb softkey from the menu. . . Figure 22 - Arbitrary Waveform 4. Select Built-in from the parameter menu and there will be a table on the display area showing the available built-in waveforms. The menu will display the waveform categories: Common, Math, Engine, etc. 5.
5. Once selected, press the knob to confirm and set the waveform or navigate to page 3/3 and press the Accept option from the menu. 6. The generator will return to the main Arb parameter menu and the waveform display area will show the waveform shape of the selected predefined arbitrary waveform. For example, the screenshot below shows the display of the predefined StairUp signal. Figure 24 - Configuring an Arbitrary Waveform 7. Adjust frequency, amplitude, and other parameters as desired.
1. Setup Press the Waveforms button. 2. Go to the second page of waveform options by pressing the 1/2 softkey. Select and press the Arb softkey from the menu. 3. Press Arb Type and select Stored Waveforms to see a list of all arbitrary waveform files created and loaded from EasyWave software. . . Figure 25 - User-Defined Arbitrary Waveform 4. Rotate the knob to choose the desired waveform. Then select Recall or press the knob to recall the corresponding waveform. 5.
6. Use the knob to navigate to the file and either press the Recall softkey or press the knob to load the waveform. If the waveform is sent to the unit using EasyWave, the file will be saved to the internal memory. If it is loaded to a USB flash drive, then the waveform can be recalled from the USB flash drive. Configure Harmonic Generator Configure the built-in harmonic generator to output harmonics with a specified order, amplitude and phase. This feature follows the Fourier series, as explained below.
Figure 27 - Harmonic Generator Item Menu Function Type Set the harmonic type to “odd”, “ever” or “all”. Order Set the order of the harmonic. Harmonic Ampl Set the amplitude of the harmonic. Harmonic Phase Set the phase of the harmonic. Return Return to the sine parameters menu. Table 6 - Harmonics Generator Menu To Select the Harmonic Type The 4050B can output odd harmonics, ever harmonics and user-defined orders of harmonics. 1. Press TYPE TO SELECT THE DESIRED HARMONIC TYPE. 2.
1. Press Order. The range is limited by the maximum output frequency of the instrument and current fundamental waveform frequency. Range: 2 to maximum output frequency of the instrument ÷ current fundamental waveform frequency The maximum is 10. 2. Use the numeric keyboard or knob to input the desired value. When the knob is used, the changes to the “Harm Order” will be immediate. When the numeric keypad is used, after entering the value, the Accept softkey should be pressed to enter the value.
AM source (internal/external), depth, modulating frequency, modulating waveform and carrier.
Figure 28 - Setting Interface of AM Modulation Function Menu Type Source Settings Description AM Amplitude modulation Internal The source is internal External The source is external. Use the [Aux In/Out] connector at the rear panel. Set the modulation depth. AM Depth Sine Square Triangle Shape UpRamp Choose the modulating waveform. DnRamp Noise Arb AM Freq Set the modulating waveform frequency. Frequency range: 1 mHz to 20 kHz (internal source only).
Selecting Modulation Source The generator can accept a modulating signal from an internal or external modulation source. Press Mod → AM → Source to select “Internal” or “External” modulation source. The default is “Internal”. Internal Source When internal AM modulation source is selected, press Shape to select Sine, Square, Triangle, UpRamp, DnRamp, Noise or Arb as the modulating waveform.
cable. 2. Select CH1 and press Mod to select the desired modulation type as well as set the corresponding parameters, and then select external modulation source. 3. Select CH2 and select the desired modulating waveform and set the corresponding parameters. 4. Press Output to enable the output of CH1. To Set Modulation Depth Modulation depth expressed as a percentage indicates the amplitude variation degree. AM modulation depth varies from 1% to 120%. Press AM Depth to set the parameter.
Figure 29 - Setting Interface of DSB-AM Modulation Function Menu Type Source Settings Description DSB-AM DSB Amplitude modulation. Internal The source is internal. External The source is external. Use the [Aux In/Out] connector at the rear panel. Set the modulating waveform frequency. Frequency range: 1 mHz to 20 kHz (internal source) DSB Freq Sine Square Triangle Shape UpRamp Choose the modulating waveform.
FM Modulation The modulated waveform consists of two parts: the carrier and the modulating waveform. In FM, the frequency of the carrier varies with the instantaneous voltage of the modulating waveform. Press Mod → Type → FM; the parameters of FM modulation are shown in Figure 30. Figure 30 - Setting Interface of FM Modulation Function Menu Type Source Settings Description FM Frequency modulation Internal The source is internal External The source is external.
Set the modulating waveform frequency. Frequency range: 1 mHz to 20 kHz (internal source). FM Freq Table 10 - FM Modulation Menu Description To Set Frequency Deviation Press FM Dev to highlight the parameter and use the numeric keyboard or arrow keys and knob to input the desired value. The deviation should be equal to or less than the carrier frequency. The sum of the deviation and the carrier frequency should be equal to or less than maximum frequency of the selected carrier waveform.
Source Internal The source is internal External The source is external. Use the [Aux In/Out] connector at the rear panel. Phase deviation ranges from 0° to 360°. Phase Dev Sine Square Triangle Shape UpRamp Choose the modulating waveform. DnRamp Noise Arb PM Freq Set the modulating waveform frequency. range:Menu 1 mHz to 20 kHz.
Figure 32 - Setting Interface of FSK Modulation Function Menu Type Source Settings Description FSK Frequency shift keying modulation Internal The source is internal External The source is external. Use the [Aux In/Out] connector at the rear panel. Set the frequency at which the output frequency shifts between the carrier frequency and the hop frequency (internal modulation only): 1 mHz to 50 kHz. Set the hop frequency.
To Set Hop Frequency The range of the hop frequency depends on the carrier frequency currently selected. Press Hop Freq to highlight the parameter and use the numeric keyboard or knob to input the desired value. Sine: 1 uHz to max output depending on model. Square: 1 uHz to max output depending on model. Ramp: 1 uHz to 500 kHz Arb: 1 uHz to 6 MHz Note: The methods of setting the parameters of FSK are similar to AM.
Type Source ASK Amplitude shift keying modulation Internal The source is internal External The source is external. Use the [Aux In/Out] connector at the rear panel. Set the frequency at which the output amplitude shifts between the carrier amplitude and zero (internal modulation only): 1 mHz to 50 kHz. Key Freq Table 13 - Table 13 - ASK Modulation Menu Description Note: The methods of setting the parameters of ASK are similar to AM.
Source Internal The source is internal External The source is external. Use the [Aux In/Out] connector at the rear panel. Set the frequency at which the output phase shifts between the carrier phase and 180° (internal modulation only): 1 mHz to 20 kHz. Key Freq Positive Polarity Set the modulating polarity Negative Table 14 - PSK Modulation Menu Description Note: The methods of setting the parameters of PSK are similar to AM.
Function Menu Type Source Settings Description PWM Pulse width modulation. The carrier is pulse. Internal The source is internal External The source is external. Use the [Aux In/Out] connector at the rear panel. Width Dev Duty Dev Set the width deviation. Set the duty deviation. Sine Square Triangle Shape UpRamp Choose the modulating waveform. DnRamp Noise Arb Set the modulating waveform frequency. PWM Freq Frequency range: 1 mHz to 20 kHz (Internal source only).
Figure 36 - Width Deviation Setting Interface The width deviation cannot exceed the current pulse width. The width deviation is limited by the minimum pulse width and current edge time setting. Duty Deviation represents the variation (%) of the modulated waveform duty cycle relative to the original duty cycle. Press Duty Dev to highlight the parameter and use the numeric keyboard or knob to input the desired value as shown in the Figure 37.
Duty deviation and width deviation are dependent. Once one parameter is changed, the other will be automatically changed. Note: The methods of setting the parameters of PWM are similar to AM. 3.5 Configure Sweep Output In the sweep mode, the generator steps from the start frequency to the stop frequency in the sweep time specified by the user. The waveforms that support sweep include sine, square, ramp and arbitrary.
Trig Out Manual Trigger a sweep manually. Off Disable trigger out. On Enable trigger out. Linear Set the sweep with linear profile. Log Set the sweep with logarithmic profile. Up Sweep upward. Down Sweep downward. Type Direction Table 16 - Sweep Function To select and configure the Sweep Function, follow the steps below: 1. Press the Sweep button. Sweep time 2. Enter the Sweep Time using the keypad or knob. Sweep Frequency 3.
a. Press the Source softkey. b. Press the Internal softkey to select an internal trigger. The generator outputs a continuous sweep waveform when internal trigger is selected. The default is “Internal”. External Trigger a. Press the Source softkey. b. Press the External softkey to select an external trigger. The generator uses the trigger signal input to the [Aux In/Out] connector at the rear panel. A sweep will be generated once a TTL pulse with specified polarity is sent.
Sweep Type 10. In the Sweep Function menu, press the Page 1/2 softkey to access the second page of Sweep Options. “Linear” and “Log” sweep types are supported. The default type is “Linear”. . A. LINEAR SWEEP In linear sweep, the output frequency of the instrument varies linearly with the sweep time. o Press the Linear softkey. A straight line is displayed on the waveform indicating that the output frequency will vary linearly. Figure 40 - Linear Sweep Interface B.
Figure 41 - Log Sweep Interface Direction 1. In the Sweep Function menu, press the Page 1/2 softkey to access the second page of Sweep Options. 2. Press the Direction softkey to toggle between the two options: Up and Down. . Up: the generator will sweep from the Start frequency to the Stop frequency. Down: the generator will sweep from the Stop frequency to the Start frequency. Note: Sweep function is only available for sine, square, ramp, and arbitrary waveforms.
Burst Type The 4050B provides three burst types: N-Cycle, Infinite and Gated. The default is N-Cycle. Burst Type, Trigger Source and Waveform Table 17 - Burst Type: Trigger Source and Waveforms Burst Type Trigger Source Waveforms N-Cycle Internal/External/Manual Sine, Square, Ramp, Pulse, Arbitrary. Infinite External/Manual Sine, Square, Ramp, Pulse, Arbitrary. Gated Internal/External Sine, Square, Ramp, Pulse, Noise, Arbitrary.
Trig Out On Enable trigger out. Cycles In N-Cycle mode, the generator will output a specified number N of waveform cycles (periods) after receiving the trigger signal. To configure Burst Cycles, follow these instructions: 1. 2. 3. 4. Press the Burst button. Press the NCycle softkey. Press the Cycles softkey. Use the numeric keyboard or knob to input the desired number of cycles. Figure 42 – Burst NCycle Infinite In infinite mode, the cycle number of the waveform is set as an infinite value.
This trigger softkey becomes available when the manual trigger is selected. Figure 43 – NCycle Burst Manual Trigger Gated In gated mode, the generator controls the waveform output according to the gated signal level. When the gated signal is ON, the generator outputs a continuous waveform. When the gated signal is OFF, the generator first completes the output of the current period and then stops. Waveforms that support gated burst include sine, square, ramp, pulse, noise and arbitrary.
Figure 44 - Negative Gated Burst To configure Burst Gated, follow the steps below: 1. Press the Burst button. 2. Press the Gated softkey. 3. Select a Polarity for the gate. This determines whether the gate will go from low to high (positive gate) or from high to low (negative gate). Common Settings for NCycle and Gated Burst: Start Phase Define the start point in a waveform. The phase varies from 0° to 360° and the default is 0°. For an Arbitrary Waveform, 0° is the first waveform point.
Burst Trigger Source The burst trigger source can be internal, external or manual. The generator will generate a burst output when a trigger signal is received and wait for the next trigger source. To configure the Trigger Source, once in the Burst menu: Internal Trigger 1. 2. 3. 4. Press the Source softkey. Press the Internal softkey. The generator outputs a continuous burst waveform when internal trigger is selected.
4 Utility Functions The utility menu can be accessed by pressing the Utility button from the front panel. This section describes all the configurable settings of this menu. Figure 45 - Utility Menu Item Description System Test/Cal Counter Set the system configuration. Test and calibrate the instrument. Frequency counter settings. Output Setup Set the output parameters of CH1 and CH2.
4.1 System Settings 1. Press the Utility button. 2. Pres the System softkey to enter the system options: Figure 46 - System Options Item Description Number format Set the numerical format. English Language Set the language. Chinese Default All the settings return to their defaults at power on. Last All the settings return to the settings at the last power down. PowerOn Set to Set all the settings to their defaults. Default On Turn on the key sound (beep). Off Turn off the key sound (beep).
1hour 2hour 5hour Off System Info Disables screensaver. View system information Firmware Update Update firmware using a USB flash drive. Help View the help information. Accept Save the current settings and return to the Utility menu. Figure 47 - System Settings Numerical Format To change the numerical format of the generator, follow these instructions: a. Press the Utility button. b. Press the System softkey. c. Press the Number Format softkey.
Figure 48 - Numerical Format Options Language Setup The generator offers two languages (English and Simplified Chinese). a. Press the Utility button. b. Press the System softkey. c. Press the Language softkey to select the desired language. Power On The 4050B series offers two options for power on settings: Last and Default. To set either option, follow these instructions: a. Press the Utility button b. Press the System softkey. c.
Set to Default Sets the unit to defaults settings: a. Press the Utility button. b. Press the System softkey. c. Press the Set to Default to set the system to the default setting.
Direction ↑ Burst Burst Period 10 ms Start Phase 0° Cycles 1 Cycle Trig Out Off Delay 521 ns Trigger Source Internal Key Sound (Beeper) Turns the instrument key press sound ON or OFF. When off, this setting will also disable the sound for any errors that may occur during front panel or remote operation. a. Press the Utility button. b. Press the System softkey. c. Press the Beeper softkey to select “On” or “Off”. The default is “On”.
Figure 49 - System Information Screen Firmware Update The software version and configuration file of the generator can be updated directly via USB flash drive. To update the firmware, follow the steps below: a. Insert USB flash drive with firmware update file (*.ADS) and configuration file (*.CFG) to USB host interface on the front panel of the generator. b. c. d. e. Press the Utility button. Press the System softkey. Press the Page 1/2 to access the second page of options.
Built-in Help The 4050B Series provides a built-in help system. There are 10 topics in the help list. Use the softkeys to navigate through the options and select a topic to read. To access the Built-in Help follow the below instructions: a. b. c. d. Press the Utility button. Press the System softkey. Press the Page 1/2 to access the second page of options. Press the Help softkey to enter the following screen: Figure 50 - Built-in Help Menu Menu Description UP Cursor upward to select.
4.2 Test/Cal This function allows the user to run self-diagnostics. a. Press the Utility button. b. Press the Test/Cal softkey. The following options will be available: Item Description ScrTest Run the screen test program. KeyTest Run the keyboard test program. LEDTest Run the key indicator lights test program. BoardTest Run the hardware circuit self-test program. Cancel Return to the Test/Cal menu. Table 23 - Self Test Menu c. To enter any test, press the corresponding sotkey: ScrTest a.
Key Test a. Press the KeyTest softkey to enter the key test interface. The on-screen white rectangle shapes represent the front panel keys. The circle between two arrows represents the knob. b. Test all keys by pressing them and knob by rotating it. Also verify that all the backlight keys illuminate correctly. The items tested will change to blue. Figure 53 - Key Test LED Test a. Press the LEDTest sotkey to enter the LED test interface. The on-screen white rectangle shapes represent the front panel keys.
Figure 54 - LED Test Board Test a. Press the BoardTest softkey to test the internal hardware of the generator. It will indicate either “pass” or “fail”. Figure 55 - Board Test 4.3 Frequency Counter The generator has a built-in frequency counter that can measure frequencies between 100 mHz to 200 MHz (DC Coupling) and 10 Hz to 200 MHz (AC Coupling). This unit has a dedicated connector in the back panel for the frequency counter. To configure the Frequency Counter follow the steps below: a.
b. Press the Counter softkey. The following options will be available: Figure 56 - Frequency Counter Function Menu State Description Off Turn the counter on. On Turn the counter off. Frequency Measure frequency. Period Measure period. PWidth Measure positive width. NWidth Measure negative width. TrigLev Set the reference frequency. The instrument will display the deviation in parts per million (ppm) between the measured frequency and the reference frequency. Set the trigger level voltage.
Period Positive pulse width (Pwidth) Negative pulse width (Nwidth) e. Press the RefFreq/TrigLevel to set the reference frequency or to set a trigger level: Reference Frequency: The counter will display the deviation in parts per million between the measured frequency and the reference frequency. Trigger Level: Sets the trigger level of the measurement system. The counter triggers and displays the measurement readings when the input signal reaches the specified trigger level.
Figure 57 - Output Setup Load For [CH1] and [CH2] connectors on the front panel, the generator has an output impedance of 50 Ω. If the load does not match the load impedance, the displayed voltage will not be the same as the output voltage setting on the generator. The Load function is used to match the displayed voltage with the expected one by scaling the display of the generator since this setting does not actually change the output impedance. Steps for setting the load: 1. Press the Utility button. 2.
Note: The Sync signal related to the waveform is not inverted when the waveform is inverted. EqPhase The unit can set both channel 1 and channel 2 to be in phase. Once the selected channel has the required phase follow the steps below to copy the phase to the other channel: 1. Press the Utility button. 2. Press the Output Setup. 3. Press the EqPhase softkey to align the phases of CH1 and CH2.
Figure 58 - Wave Combine Menu Function Menu CH1 Switch CH2 Switch Description CH1 Output the CH1 waveform. CH1+CH2 Output the CH1+CH2 waveform. CH2 Output the CH2 waveform. CH1+CH2 Output the CH1+CH2 waveform. Return Notes: Save the current settings and exit the current interface. 1. Combining with a Square waveform is not possible. 2. When the waveform combining function is enabled, the output impedance of the two channels will be set to be the same as the currently operated channel. 4.
Function Menu Description CH1=>CH2 Copy all parameters and states of CH1 to CH2. CH2=>CH1 Copy all parameters and states of CH2 to CH1. Accept Perform the current selection and return to the Utility menu. Cancel Give up the current selection and return to the Utility menu. Table 25 - Channel Copy Options To configure channel copy options, follow the steps below: 1. Press the Utility button. 2. Press the CH Copy Coupling softkey. 3. Press the Channel Copy softkey to enter the following interface.
of the other channel will be changed automatically and always keeps the specified frequency deviation/ratio, amplitude deviation/ratio or phase deviation/ratio relative to the modified channel. To configure channel coupling settings, follow the instructions below: 1. Press the Utility button. 2. Press the CH Copy Coupling softkey. 3.
2. Press AmplMode to choose “Deviation” or “Ratio” and use the numeric keyboard or knob to input the desired value. Deviation: the amplitude deviation between CH1 and CH2. The resulting signal is represented by: AmplCH2-AmplCH1 = AmplDev. Ratio: the amplitude ratio of CH1 and CH2. The resulting signal is represented by: AmplCH2/AmplCH1 = AmplRatio. Phase Coupling To configure the Phase Coupling function, follow the steps below: 1. Press PhaseCoup to turn phase coupling “On” or “Off”.
Figure 61 - Tracking 4. Press PhaseDev to enter the following menu: Figure 62 - Phase Deviation 5. Then use the numeric keyboard or knob and arrow keys to input the desired value for the phase deviation between CH1 and CH2. The resulting signal is represented by: PhaseCH2-PhaseCH1=PhaseDev.
4.6 Remote Interface The 4050B Series can be controlled remotely via USB, LAN and GPIB (optional) interfaces. 1. Press the Utility button. 2. Press the Page 1/2 softkey. 3. Press the Interface softkey to set the LAN parameters or GPIB address in the following screen: Figure 63 - Remote Interface The unit will recognize if USBTMC or GPIB interface is being used and output to that interface. GPIB takes precedence as it locks the unit to GPIB mode when the adapter is connected to USB host in the front panel.
USB Interface The instrument has a USB interface (USBTMC) in the rear panel for remote communication with EasyWave software. Note: A type A to type B USB cable is required for PC connectivity. To connect with EasyWave software, you must install NI-VISA. NI-VISA is downloadable from LabVIEW™ website. There are no settings for USBTMC. Make sure to set the LAN State to OFF. Note: Users who have LabVIEW™ or NI-VISA installed will automatically have this driver in their system.
Power indicator LED GPIB USB PC Generator AK40G Communication indicator LED Note: Once connected, do not unplug the adapter on either end before powering down the instrument and the computer first. Configure GPIB address 1. 2. 3. 4. Press the Utility button. Press the 1/2 softkey to enter the second page of the Utility menu. Press the Interface softkey to select the remote interface. Press GPIB and use the rotary knob to set the GPIB address of the generator. The address range is 1 to 30. 5.
Note: EasyWave software does not support GPIB interface. It must be used with USB interface only. Note: All supported remote commands are described in the programming manual which can be downloaded from the B&K Precision website www.bkprecision.com. LAN The 4050B SERIES can communicate with a PC through a LAN interface. Users can view and modify the LAN parameters. 1. Connect the generator to your local area network using the network cable. 2. Configure network parameters: a. b. c. d. e.
Press IP Address and use the arrow keys and numeric keyboard or knob to enter the IP address. The setting is stored in non-volatile memory and will be loaded automatically when the generator is powered on. To Set Subnet Mask The format of subnet mask is NNN. NNN. NNN.NNN and each NNN ranges from 0 to 255. Press Subnet Mask and use the numeric keyboard or knob to enter your desired subnet mask. The setting is stored in non-volatile memory and will be loaded automatically when the generator is powered on.
4.7 Sync Output The generator provides Sync output through the [Aux In/Out] connector on the rear panel. When the synchronization is on, the port can output a CMOS signal with the same frequency as basic waveforms (except Noise and DC), arbitrary waveforms, and modulated waveforms (except external modulation). Either channel 1 or channel 2 can be the waveform this signal is synchronized with. Figure 66 - Sync Output Function Description Menu State Off Turn off the sync output.
Syncing Signals of Different Waveforms Basic Waveform and Arbitrary Waveform 1. When the frequency of the waveform is less than or equal to 10 MHz, the sync signal is a Pulse with approximately 50 ns pulse width and the same frequency as the waveform. 2. When the frequency of the waveform is greater than 10 MHz, there is no sync signal output. 3. Noise and DC waveforms will have no sync signal output. Modulated Waveform 4.
prompt message “No external clock source!” would be displayed and the clock source would be switched to “Internal”. 4. The unit can also output a 10 MHz clock signal when the 10MOut is enabled. If the output of the clock signal is enabled, then the [10MHz In/Out] connector in the rear panel will function as an clock signal output only.
Figure 67 - Phase-Locked Mode Independent Mode When changing the frequency, both channels’ DDS don’t reset and the phase deviation between CH1 and CH2 changes at random. When the independent mode is enabled, the phase parameter cannot be modified, and the Phase menu will be hidden as shown below.
4.10 Over Voltage Protection 1. Press the Utility button. 2. Press the Page 1/2 softkey. 3. Press the OverVoltage Protection to turn this function on or off. When overvoltage protection occurs, a message will be displayed and the output will be disabled. If set to ON, overvoltage protection of CH1 and CH2 will take effect when either of the following conditions are met. The absolute value of input voltage is higher than 11 V ± 0.
5 Store and Recall The 4050B series can store the current instrument state and user-defined arbitrary waveform data in internal or external memory and recall them when needed. Press Store/Recall to enter the following interface.
File Type State Settings files. Data Arbitrary waveform file. Browse View the current directory. Save Save the waveform to the specified path. Delete Recall the waveform or setting information in the specific position of the memory. Delete the selected file. Copy Copy the selected file. Paste Paste the selected file. Cancel Exit the Store/Recall interface. Recall Figure 71 - Store/Recall Interface 5.
Figure 72 - Storage System Note: The 4050B series can only identify files of which filenames consist of English letters, numbers and underscore. If other characters are used, the name may be displayed in the store and recall interface with errors or not found at all. Browse Use the knob to shift between the directories to choose Local (C:) or USB Device (0:). Choose Browse, press the knob to open the current directory. Use the knob to switch between folder and files under the current directory.
Data File The 4050B can recall the data files in “*.csv” or “*.dat” format from the external memory and transfer them into “*.bin” format then store them in the internal memory. When it is done, the generator will enter the arbitrary waveform interface automatically. In addition, users can edit arbitrary waveforms with PC software — EasyWave, download them to the internal memory through remote interface and store them (“*.bin” format) in the internal memory. To download EasyWave, please visit: www.
Figure 73 - Filename Input Interface Item Description Up Cursor upward to select. Down Cursor downward to select. Select Select the current character. Delete Delete the current character. Save Store the file with the current name. Cancel Return to the store/Recall interface. a. Users can select the desired letter, number or character from the virtual soft keyboard by using the knob or Up and Down menus. b. Then choose Select to display the character selected in the filename area. c.
5.4 Recall State File or Data File To recall the instrument state or arbitrary waveform data, the procedures are as follows: 1. Choose the file type. Press Store/Recall → File Type, and choose state or data as the storage type. 2. Choose the file to be recalled. Rotate the knob to select the file you want to recall. 3. Recall the file. Choose Recall, press the knob, the generator will recall the selected file and display corresponding prompt message when the file is read successfully. 5.
2. Choose the file to be copied. Rotate the knob to select USB Device (0:) and press the knob to open its directory. Then rotate the knob to select the file you want to copy and press Page 1/2 → Copy. 3. Paste the file. Rotate the knob to select Local (C:) and press the knob to open its directory. Then press Paste.
6 Examples The examples below use the default settings of the instrument except in special cases.
Example 1: Generate a Sine Waveform Generate a sine waveform with 1 MHz frequency, 5 Vpp amplitude and 1 V DC offset. Set the frequency. 1. Press Waveforms → Sine →Frequency/Period and choose Frequency, which will display in a blue color. 2. Input ‘1’ from the keyboard and choose the unit ‘MHz’. The frequency will be set to 1 MHz. Set the Amplitude. 1. Press Amplitude/HighLevel to choose Amplitude which will display in a blue color. 2. Input ‘5’ from the keyboard and choose the unit ‘Vpp’.
Example 2: Generate a Square Waveform Generate a square waveform with 5kHz frequency, 2 Vpp amplitude, 1 V DC offset and 30% duty cycle. Set the frequency. 1. Press Waveforms → Square → Frequency/Period and choose Frequency, which will display in a blue color. 2. Input ‘5’ from the keyboard and choose the unit ‘kHz’. The frequency is set to 5 kHz. Set the Amplitude. 1. Press Amplitude/HighLevel to choose Amplitude which will display in a blue color. 2.
Example 3: Generate a Ramp Waveform Generate a ramp waveform with 10 μs period, 100 mVpp amplitude, 20 mV DC offset, 45° phase and 30% symmetry. Set the Period. 1. Press Waveforms → Ramp → Frequency/Period and choose Period which will display in a blue color. 2. Input ‘10’ from the keyboard and choose the unit ‘μs’. The period is set to 10 μs. Set the Amplitude. 1. Press Amplitude/HighLevel to choose Amplitude which will display in a blue color. 2.
Figure 76 - Generate a Ramp Waveform Example 4: Generate a Pulse Waveform Generate a pulse waveform with 5 kHz frequency, 5 V high level, -1 V low level, 40 μs pulse width and 20 ns delay. Set the Frequency. 1. Press Waveforms → Pulse → Frequency/Period and choose Frequency, which will display in a blue color. 2. Input ‘5’ from the keyboard and choose the unit ‘kHz’. The frequency is set to 5 kHz. Set the HighLevel. 1.
1. Press Delay to choose Delay which will display in a blue color. 2. Input ‘20’ from the keyboard and choose the unit ‘ns’. The delay is set to 20 ns. When the frequency, high level, low level, pulse width and delay are set, the waveform generated is shown in Figure 77. Figure 77 - Generate a Pulse Waveform Example 5: Generate a Noise Generate a noise waveform with 0.5 V standard deviation and a 1 V mean. Set the Stdev. 1.
Figure 78 - Generate a Noise Waveform Example 6: Generate a DC Waveform Generate a 3 V DC waveform. Choose the DC waveform. 1. Press Waveforms →Page 1/2→ DC to choose the DC waveform. Set the Offset. 1. Press Offset and choose Offset which will display in a blue color. 2. Input ‘3’ from the keyboard and choose the unit ‘Vdc’. The DC offset is set to 3 V. When the DC offset is set, the waveform generated is shown in Figure 79.
Figure 79 - Generate a DC Waveform Example 7: Generate a Linear Sweep Waveform Generate a sine sweep waveform whose frequency starts at 100 Hz and sweeps to a frequency of 10 kHz. Use internal trigger mode, linear sweep, and a sweep time of 2 s. Set the sweep function. Press Waveforms and choose the sine waveform as the sweep function. The default setting of the source is internal. Set the amplitude and offset. 1. Press Amplitude/HighLevel to choose Amplitude which will display in a blue color.
Set the stop frequency. Press StopFreq, input ‘10’ from the keyboard and choose the unit ‘kHz’ to set the stop frequency to 10 kHz. Set the sweep profiles. Press Type and choose Linear. When all parameters above are set, the linear sweep waveform generated is shown in Figure 80. Figure 80 - Generate a Linear Sweep Waveform Example 8: Generate a Burst Waveform Generate a burst sine waveform with 5 cycles. The burst period is 3 ms. Use internal trigger and 0° start phase. Set the burst function.
from the keyboard and choose the unit ‘Vpp’ to set the amplitude to 4 Vpp. 3. Press Offset/LowLevel to choose Offset which will display in blue color. Input ‘0’ from the keyboard and choose the unit ‘Vdc’ to set the DC offset to 0 V DC. Set the burst mode. Press Burst → NCycle and choose N-Cycle Mode. The default setting of the source is internal. Set the burst period. Press Burst Period, input ‘3’ from the keyboard and choose the unit ‘ms’ to set the burst period to 3 ms. Set the start phase.
Example 9: Generate an AM Modulation Waveform Generate an amplitude-modulated waveform with 80% depth. The carrier is a sine wave with 10 kHz frequency and the modulating wave is a sine wave with 200 Hz frequency. Set the frequency, amplitude and offset of the carrier wave. 1. Press Waveforms and choose the sine waveform as the carrier wave. 2. Press Frequency/Period and choose Frequency which will display in a blue color.
Example 10: Generate an FM Modulation Waveform Generate an FM modulation waveform: the carrier is a sine wave with 10 kHz frequency, the modulating wave is a sine wave with 1 Hz frequency and 2 kHz frequency deviation. Set the frequency, amplitude and offset of the carrier wave. 1. Press Waveforms and choose the sine waveform as the carrier wave. 2. Press Frequency/Period and choose Frequency which will display in a blue color.
Example 11: Generate a PM Modulated Waveform Generate a PM modulated waveform, the carrier is a sine wave with 10 kHz frequency, and the modulating wave is a sine wave with 2 kHz frequency and 90° phase deviation. Set the frequency, amplitude and offset of the carrier wave. 1. Press Waveforms, and choose the sine waveform as the carrier wave. 2. Press Frequency/Period and choose Frequency which will display in a blue color.
Example 12: Generate an FSK Modulated Waveform Generate an FSK modulated waveform with a 200 Hz key frequency. The carrier is a sine wave with a 10 kHz frequency and the hop frequency is 500 Hz. Set the frequency, amplitude and offset of the carrier wave. 1. Press Waveforms and choose the sine waveform as the carrier wave. 2. Press Frequency/Period and choose Frequency which will display in a blue color. Input ’10’ from the keyboard and choose the unit ‘kHz’ to set the frequency to 10 kHz. 3.
Example 13: Generate an ASK Modulation Waveform Generate an ASK modulation waveform with 500 Hz key frequency. The carrier is a sine wave with 5 kHz frequency. Set the frequency, amplitude and offset of the carrier wave. 1. Press Waveforms and choose the sine waveform as the carrier wave 2. Press Frequency/Period and choose Frequency which will display in a blue color. Input ‘5’ from the keyboard and choose the unit ‘kHz’ to set the frequency to 5 kHz 3.
Example 14: Generate a PSK Modulated Waveform Generate a PSK modulated waveform with 200 Hz key frequency. The carrier is a sine wave with 1 kHz frequency. Set the frequency, amplitude and offset of the carrier wave. 1. Press Waveforms and choose the sine waveform as the carrier wave 2. Press Frequency/Period and choose Frequency which will display in a blue color. Input ‘1’ from the keyboard and choose the unit ‘kHz’ to set the frequency to 1 kHz 3.
Example 15: Generate a PWM Modulated Waveform Generate a PWM modulated waveform with 200 Hz key frequency. The carrier is a pulse wave with 5 kHz frequency. Set the frequency, amplitude and offset of the carrier wave. 1. Press Waveforms and choose the Pulse waveform as the carrier wave 2. Press Frequency/Period and choose Frequency which will display in a blue color. Input ‘5’ from the keyboard and choose the unit ‘kHz’ to set the frequency to 5 kHz 3.
Example 16: Generate a DSB-AM Modulated Waveform Generate a DSB-AM modulated waveform with 100 Hz modulating frequency. The carrier is a sine wave with 2 kHz frequency. Set the frequency, amplitude and offset of the carrier wave. 1. Press Waveforms and choose the sine waveform as the carrier wave. 2. Press Frequency/Period and choose Frequency which will display in a blue color. Input ‘2’ from the keyboard and choose the unit ‘kHz’ to set the frequency to 2 kHz 3.
Troubleshooting Guide Below are some frequently asked questions and answers. Please check if any apply to your power supply before contacting B&K Precision. Q: I cannot power up the generator o Check that the power cord is securely connected to the AC input and there is live power from your electrical AC outlet. o Verify that the AC power coming from the mains is the correct voltage. The generator can accept a specific range of AC input voltages. Refer to section “2.1”.
SPECIFICATIONS Model 4053B 4054B Channels 4055B 2 Frequency Characteristics Sine & Square 1 µHz – 10 MHz 1 µHz – 30 MHz Triangle, Ramp 1 µHz – 500 kHz Pulse 1 µHz – 12.5 MHz Noise (-3 dB) > 60 MHz Arbitrary 1 µHz – 6 MHz Accuracy ± 25 ppm (1 year) Resolution 1 µHz 1 µHz – 60 MHz Arbitrary Characteristics Built-in Waveforms 196 built-in waveforms (includes DC) Waveform Length 16k points / Ch Vertical Resolution 14 bits Sampling Rate 150 MSa/s Minimum Rise/Fall Time 6.
Total Harmonic Distortion (Sine) Spurious (non-harmonic) 10 Hz – 20 kHz at 0 dBm, < 0.075% DC – 10 MHz, < -65 dBc / 10 MHz – 30 MHz, < -55 / 30 MHz – 60 MHz, < -40 (0 dBm input signal) Rise/Fall Time (square) Variable Duty Cycle (square) < 4.2 ns (10 % – 90 %, at 1Vpp into 50 Ω) 0.001% - 99.999% (depending on frequency setting) Asymmetry (50% duty cycle) 1% of period + 20 s (typical,1 kHz, 1 Vpp) Jitter (rms) cycle to cycle (square) 300 ps + 0.
PM Phase Deviation 0 – 360 °, 0.
Frequency Counter Measurement Frequency, period, duty cycle, positive/negative pulse width Measurement Range 100 mHz – 200 MHz (DC coupling) 10 Hz – 200 MHz (AC coupling) Input Range 100 mV to ± 2.5 V (< 100 MHz, DC coupling) 200 mV to ± 2.
SERVICE INFORMATION Warranty Service: Please go to the support and service section on our website at www.bkprecision.com to obtain a RMA #. Return the product in the original packaging with proof of purchase to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Non-Warranty Service: Please go to the support and service section on our website at www.bkprecision.com to obtain a RMA #.
LIMITED THREE-YEAR WARRANTY B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of three years from date of purchase. B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt.
22820 Savi Ranch Parkway Yorba Linda, CA 92887 www.bkprecision.com © 2017 B&K Precision Corp.