User Manual
96 
Figure 50 Comparator Display Percent Error 
  ABS (Absolute): The user can set absolute HIGH and LOW limits. The instrument will 
compare the measured value to the absolute limits and determine if the DUT is HI 
(above the upper limit), LO (below the lower limit), or IN (within limits). 
3.3  Bin Display 
The Bin Display (BIN DISP) page places the resistance measurement in up to 10 user-defined 
bins. See 0  for instructions to define bins. The bin operation is executed when the instrument is 
triggered. To access the page press the DISP button and select BIN DISP in menu options. 
Figure 51 - Bin Display 
When the measurement is outside of a bin’s tolerance, it is defined as NOT GOOD (NG). When 
the measurement is inside of a bin’s tolerance, it is defined as GOOD (GD). 