User Manual
95
The following parameters are accessible from the Comparator Display.
3.2.1 OPTIONS
The OPTIONS key allows the user to access the additional comparator option menu.
The options are:
DISP (ON/OFF): Toggle the measurement results display on and off
COMP BEEP (OFF/NG/GD): Toggles beep between off, not good (NG) and good (GD).
COUNT (ON/OFF): ON begins the comparator count if COMP is ON. Count increments
based on trigger settings. OFF stops the count.
COUNT CLEAR: HI, LO, IN, and TOT counts are reset to 0.
SAVE DATA: Enabled when a USB flash drive is attached. When activated, resistance and
temperature data is continuously saved to an auto-generated CSV onto the drive. The
key changes to STOP SAVE to terminate the function. SAVE DATA is disabled if no USB
flash drive attached.
Removing the USB flash drive before pressing STOP SAVE will result in the loss of the
recorded data. The CSV is rewritten every time.
3.2.2 COMP
The COMP key turns the comparator functions ON and OFF. Results are in the results display
area. COMP: NC will be displayed if COMP is OFF.
3.2.3 COMP MODE
The COMP MODE key allows the user to access two comparator modes.
% (Percent Error): The user can set the nominal value (NOM) and the percent error (%).
If the nominal value is 100 and percentage is 10, the tolerance will be 100±10%. The
instrument will compare the measured value to the tolerance and determine if the DUT
is HI (above tolerance), LO (below tolerance), or IN (within tolerance). The display will
show ± percent (∆) difference the measurement is from the nominal value.
OPTIONS
• DISP
• COMP BEEP
• COUNT
• COUNT CLEAR
• SAVE DATA
COMP
• OFF
• ON
COMP MODE
• %
• NOM
• %
• ABS
• HIGH
• LOW