User Manual
93
AUTO: Automatically selects the range mode depending on the resistance
detected.
HOLD: Lock to the current resistance measurement range.
↑ (+): Increase measurement range and sets the measurement to HOLD.
↓ (- ): Decrease measurement range and sets the measurement to HOLD.
If FUNC is set to T, resistance RANGE menu options are disabled and will display AUTO.
The temperature measurement (T) range is fixed at -10
o
C to 99.9
o
C.
3.1.4 SPEED
The SPEED controls the measurement time. In SLOW1 and SLOW2 mode noise is lower but
measurement time is increased. FAST mode performs high speed measurements but with
increased noise. See 1.1 for timing details.
Follow the instructions below to set the measurement speed.
1. Touch the SPEED key, the following menu options will be displayed:
FAST
MED
SLOW1
SLOW2
3.1.5 TC/∆t
The TC/∆t option allows the user to enable the temperature compensation functions. Touch
the TC/∆t key, the following menu options will be displayed:
OFF: Disable all temperature compensation functions
TC: Enable Temperature Correction. See 4.2.4 for details.
∆t: Enable Temperature Conversion. See 4.2.5 for details.
Note: In SLOW2, SLOW1 or MED mode, the resolution of the measurement is 6 digits. FAST mode is 5 digits.