6  Specifications 
All specifications apply to the unit after: 
Temperature Stabilization time: 30 mins 
Operating Temperature: 23 °C ± 5 °C 
Relativity Humidity: ≤ 80% 
Specifications are subject to change without notice. 
Rd = Measured Value 
Fs = Full range 
Resistance Measurement 
Reading Digits 
4 ½ digits 
Measurement Range 
1 mΩ to 20 kΩ 
Resistance Range 
Test Current 
Resolution 
Accuracy 
Rd % + digits 
Open Circuit 
Voltage 
20 mΩ 
1 A 
1 μΩ 
0.1 + 3 
0.7 V 
200 mΩ 
100 mA 
10 μΩ 
2 Ω 
100 mA 
100 μΩ 
0.1 + 2 
20 Ω 
10 mA 
1 mΩ 
3 V 
200 Ω 
1 mA 
10 mΩ 
2 kΩ 
100 μA 
100 mΩ 
20 kΩ 
100 μA 
1 Ω 
LPR Resistance Measurement 
2 Ω 
10 mA 
100 μΩ 
0.2 + 5 
40 mV 
20 Ω 
1 mA 
1 mΩ 
200 Ω 
100 μA 
10 mΩ 
2 kΩ 
10 μA 
100 mΩ 
Accuracy = (measurement value * Rd%) + (least significant digit) 
Example calculation: 
Measurement = 10.404 Ω; RANGE = 20 Ω 
Accuracy = (10.404 Ω * 0.1%) + 0.002 = 0.012404 Ω 
Result: 10.404 Ω ± 0.012 Ω 