The following parameters are accessible from the Measurement Setup page: 
Figure 25 - MEAS SETUP Menu Tree 
4.1.1  FUNC 
This function is identical to FUNC on <MEAS DISP> page. See 3.1.2 for details. 
4.1.2  RANGE  
This function is identical to RANGE on <MEAS DISP> page. See 3.1.3 for details. 
4.1.3  TRIG 
Press TRIG to access the 4 trigger mode options: 
  INT: (Internal), continuously tests the DUT and displays the result, (default mode). 
  MAN: (Manual), press the TRIGGER button to test a DUT once and display the result. 
  EXT: (External), trigger the instrument through the  handler pin 1. 
  BUS: Trigger the instrument through the COM interface with a software command. 
Note: The instrument is in waiting mode until triggered in all trigger modes except INT 
4.1.4  TRIG DELAY 
The TRIG DELAY enables the user to select between AUTO and MANU (manual) modes. 
Automatic trigger delay is a maximum of 30 ms. To set a manual trigger delay: 
1.  Press MANU  
2.  Press INPUT to open the numeric keypad. 