User Manual
26
Connect AC power cord to the AC receptacle in the rear panel and press the power switch to
the | ( ) ON position to turn ON the instrument. The instrument will have a boot screen
while loading (see Figure 10), after which the main screen will be displayed.
After power on, the loading screen will be displayed for about five seconds. The firmware
version is found on the loading screen:
Figure 10 - Boot Screen
If password protection is enabled, the user will have to input the password to operate the unit.
The default password is 2840. See 4.3.4 Password for more details.
2.5 Cable Calibration
The resistance in the 4-terminal cables can be compensated for through the zero adjustment (0
ADJ) function. This should be performed when connecting a new cable to the instrument, when
the environment temperature changes ±5 °C, or when desired.
Complete the following steps to perform the cable calibration:
1. Warm-up the instrument for 30 minutes.
2. Short the 4-terminal connectors together (see Figure 11).
3. Ensure DRIVE+ is aligned with DRIVE- and SENSE+ is aligned with SENSE-.
4. Select a resistance range or AUTO to calibrate for all ranges.
5. Initiate the calibration function. Press 0 ADJ on the front panel.
6. Calibration time is affected by the SPEED, AVERAGE samples, and auto ranging.