User Manual
131 
8.2  Output Signal 
ERR: Error signal output is divided into synchronous measurement error signal output and 
asynchronous measurement error signal output. The measurement error signal is outputted 
together with EOC as synchronous error signal output, not asynchronous error signal output. 
INDEX: Transitions from low to high to indicate that the instrument measurement is finished. 
The instrument begins calculating the DUT’s resistance after the measurement is finished. 
EOC: End of conversion signal. Transitions from high to low when measurement calculations 
begin. Transitions from low to high when measurement calculations end. 
Hi, IN, Lo: Comparator results. 
BIN0 to BIN9, OB (out of bins): If the DUT value conforms to the bin parameters, the 
corresponding bin no. output will be low. If no bins are satisfied, the OB signal will transition 
from high to low. 
OUT0 to OUT7: Output signal controlled by :IO:OUT command. OUT0 is low-order and OUT7 is 
high-order. For example, if IO: OUT 52 the output is 
Value 
52(decimal) 
OUTX 
OUT7 
OUT6 
OUT5 
OUT4 
OUT3 
OUT2 
OUT1 
OUT0 
Bit 
0 
0 
1 
1 
0 
1 
0 
0 
BCD1-0 to BCD6-3: Binary coded decimal for the DUT’s 6 digit measurement value. BCDx-0 is 
low-order and BCDx-3 is high-order. BCD1-x is low-order and BCD6-x is high-order. For example, 
if the measurement resistance is 498.992 the output is 
Value 
4 
9 
8 
9 
9 
2 
BCDx-x 
BCD6- 
BCD5- 
BCD4- 
BCD3- 
BCD2- 
BCD1- 
3 
2 
3 
2 
3 
2 
1 
0 
3 
2 
1 
0 
3 
2 
1 
0 
3 
2 
1 
0 
3 
2 
1 
0 
Bit 
0 
1 
0 
0 
1 
0 
0 
1 
1 
0 
0 
0 
1 
0 
0 
1 
1 
0 
0 
1 
0 
0 
1 
0 
8.3  Power and Ground 
COM: External power supply EXTV reference ground 
EXTV: External power supply (5 to 24 VDC) 
INT.GND: Internal and INT.VCC ground 
INT.VCC: Internal power supply outputs 5 VDC 
Note: When using internal power supply, COM and INT.GND can be shorted and EXTV and INT.VCC can be 
shorted. Do not short when an external power supply is used. 










