User Manual
124 
Resistance Measurement Delay 
Rang
e (Ω) 
20m 
200m 
2 
20 
200 
2k 
20k 
100k 
1M 
10M 
100M 
FUNC 
R 
(ms) 
OVC 
OFF 
30 
30 
3 
3 
3 
3 
3 
10 
50 
100 
1000 
OVC 
ON 
100 
100 
100 
100 
100 
100 
100 
― 
― 
― 
― 
FUNC 
LPR 
(ms) 
OVC 
OFF 
― 
― 
3 
3 
3 
15 
― 
― 
― 
― 
― 
OVC 
ON 
― 
― 
100 
100 
100 
100 
― 
― 
― 
― 
― 
6.5  Accuracy for Temperature Measurement (TPTC2) 
Temperature range 
-10.0 to 39.9 °C 
40.0 to 99.9 °C 
Resolution 
0.1 °C 
0.1 °C 
Accuracy in six months 
±0.30%Rd ± 0.5 °C * 
±0.30%Rd ± 1.0 °C 
Accuracy in one year 
±0.45%Rd ± 0.8 °C 
±0.45%Rd ± 1.5 °C 
*Accuracy=0.3% x measured value ± 0.5 °C
6.6  Accuracy for Temperature Measurement (Analog Input) 
Analog Input 
Input voltage range 
0 to 2 V 
Temperature range display 
-99.9 °C to 999.9 °C 
Resolution 
1 mV 
Accuracy 
±1% T
R
 ± 3 mV 
Accuracy= 1%* (
 – 
) + 0.3%*(
 
) 
 
:
The temperature measured under input voltage of 1V. 
 
:
The temperature measured under input voltage of 0V. 
 
:
The current measured temperature. 