User Manual

124
Resistance Measurement Delay
Rang
e (Ω)
20m
200m
2
20
200
2k
20k
100k
1M
10M
100M
FUNC
R
(ms)
OVC
OFF
30
30
3
3
3
3
3
10
50
100
1000
OVC
ON
100
100
100
100
100
100
100
FUNC
LPR
(ms)
OVC
OFF
3
3
3
15
OVC
ON
100
100
100
100
6.5 Accuracy for Temperature Measurement (TPTC2)
Temperature range
-10.0 to 39.9 °C
40.0 to 99.9 °C
Resolution
0.1 °C
0.1 °C
Accuracy in six months
±0.30%Rd ± 0.5 °C *
±0.30%Rd ± 1.0 °C
Accuracy in one year
±0.45%Rd ± 0.8 °C
±0.45%Rd ± 1.5 °C
*Accuracy=0.3% x measured value ± 0.5 °C
6.6 Accuracy for Temperature Measurement (Analog Input)
Analog Input
Input voltage range
0 to 2 V
Temperature range display
-99.9 °C to 999.9 °C
Resolution
1 mV
Accuracy
±1% T
R
± 3 mV
Accuracy= 1%* (

) + 0.3%*(


)

The temperature measured under input voltage of 1V.

The temperature measured under input voltage of 0V.
The current measured temperature.