User Manual
121 
Display Digits 
SLOW2, SLOW1, MED 
FAST 
Display Digits 
6 digits 
5 digits 
Temperature range 
3 digits, maximum display 
number: 999.9 °C 
3 digits, maximum display 
number: 999.9 °C 
Measurement Function 
Resistance 
Measurement Time 
FAST: 7 ms, MED: 22 ms, SLOW1: 102 ms, SLOW2: 402 ms 
(+20 ms when DISPLAY is ON) 
Temperature 
Measurement Time 
100 ± 10 ms 
Low Voltage 
Measurement (LPR) 
Open voltage ≤ 60 mV 
Effective range: 2 Ω, 20 Ω, 200 Ω, 2 kΩ 
Average Samples 
1 to 255 
Temperature Measurement 
TPTC2 Sensors Range 
-10.0  to 99.9  
Analog Input 
Input Voltage Range: 0 V to 2 V 
Temperature Range:-99.9  to 999.9  
General specification 
External USB memory 
Save / recall configuration files, screenshots, and measurement logs 
Remote interface 
USB (USBTMC or virtual COM), RS232, LAN, HANDLER 
Display 
24-bit, 400 x 272 touch TFT LCD screen 
Operating 
Temperature, 
Humidity 
0 °C to 40 °C, ≤ 90%RH 
Temperature, 
Humidity 
for basic accuracy 
23 ± 5 °C, ≤ 80%RH 
AC 
Input 
Voltage 
110 V to 240 V (±10%) 
Frequency 
50 Hz to 60 Hz (±5%) 
Power Consumption 
≤ 30 VA 
Dimensions 
(WxHxD) 
255 mm x 110 mm x 361 mm 
(10.04 in x 4.33 in x 14.22 in) (with bezel) 
Weight 
Approx. 3.6 kg 
6.2  Basic Accuracy for Resistance Measurement 
Range 
Max 
display 
value 
OVC 
± (Rd% + Fs%) 
Current 
Range 
Open 
Circuit 
Voltage 
SLOW2 
SLOW1 
MED 
FAST 
20 mΩ 
20.0000 
± 0.2000 
OFF 
0.25+0.015 
0.25+0.017 
0.25+0.02 
0.25+0.025 
1A ±5% 
5Vmax 
ON 
0.25+0.001 
0.25+0.001 
0.25+0.001 
0.25+0.004 