User Manual
104 
Note: The instrument is in waiting mode until triggered. 
4.1.5  TRIG DELAY 
The  TRIG  DELAY  enables  the  user  to  select  between  AUTO  and  MANU  (manual)  modes. 
Automatic trigger delay is determined by Table 11 Auto Trigger Delay. To set a manual trigger 
delay: 
1.  Press MANU 
2.  Press INPUT for open a numeric keypad. 
3.  Set delay. Manual trigger delay range is: 0 ms to 9999 ms. 
Note: TRIG DELAY cannot be less than the DETECT time. The instrument will prompt a message “Data out of 
range.” 
Range(Ω) 
20m 
200m 
2 
20 
200 
2k 
20k 
100k 
1M 
10M 
100M 
Resistance 
Measurement 
Delay (ms) 
OVC OFF 
30 
30 
3 
3 
3 
3 
3 
10 
50 
100 
1000 
OVC ON 
100 
100 
100 
100 
100 
100 
100 
― 
― 
― 
― 
Resistance 
Measurement at 
low voltage 
Delay(ms) 
OVC OFF 
― 
― 
3 
3 
3 
15 
― 
― 
― 
― 
― 
OVC ON 
― 
― 
100 
100 
100 
100 
― 
― 
― 
― 
― 
Table 11 Auto Trigger Delay 
4.1.6  AVERAGE 
Press AVERAGE  to  open a numeric keypad. The user  can  set the  average number  a samples 
taken per trigger, ranging from 1 to 255. The larger the sample sizes reduces variation at the 
cost of a longer time to display the result. 
4.1.7  200mΩ 
The 200 mΩ resistance measurement range has two current source options. 
  1A: Open voltage is 5 V max. 
  100mA: Open voltage is 2.6 V max. 
4.1.8  DETECT 
DUT detection is  the time between trigger initiation  and when  the measurement  starts.  The 
user can select from the following options: 
  AUTO: Internally set timing. 
Note: If delay time is set to 0 ms, the contact settling time detection cannot be executed. It is recommended to 
set the delay time to be greater than 1 ms. 