The following parameters are accessible from the Measurement Setup page:
4.1.1 mR + b
This algebraic function displays a modified result to the resistance measurement display. The
function can be selected ON or OFF. The variables m and b open a numeric keypad to enter a
value.
Example calculation: R (measurement) = 10.0404 Ω; m = 2; b = 5
m * R + b = 2 * 10.0404 Ω + 5 = 25.0808 Ω
25.0808 Ω will be displayed in the measurement result area instead of 10.0404 Ω.
4.1.2 FUNC
This function is identical to FUNC on <MEAS DISP> page. See 3.1.2 for details.
4.1.3 RANGE
This function is identical to RANGE on <MEAS DISP> page. See 3.1.3 for details.
4.1.4 TRIG
Press TRIG to access the 4 trigger mode options:
INT: Continuously test a DUT and display the result.
MAN: Press TRIGGER to test a DUT once and display the result.
EXT: Trigger the instrument through the handler pin 4.
BUS: Trigger the instrument through the COM interface.
mR+b
•OFF
•ON
FUNC
•R
•R-T
•T
•LPR
•LPR-T
RANGE
•AUTO
•HOLD
•↑
•↓
TRIG
•INT
•MAN
•EXT
•BUS
TRIG DELAY
•AUTO
•MANU
•INPUT
AVERAGE
•1 - 255
200 mΩ
•1A
•100mA
DETECT
•AUTO
•MANU
•INPUT
Meas
Mode
•SLOW
•FAST
SPEED
•FASST
•MED
•SLOW1
•SLOW2
m
•-1000000
to
1000000
b
•-1000000
to
1000000
Adjust
•AUTO