User Manual

Chapter 3
FUNCtion
3.1 FUNCtion:IMPedance
Description Set or query the measurement mode.
Command Syntax FUNCtion:IMPedance <STR>
Parameters R resistance
RT resistance and temperature 2841 Only
T temperature 2841 Only
LPR low power resistance
LPRT low power resistance and temperature 2841 Only
Example Set to normal power resistance and temperature measurement.
func:imp rt
ery Syntax FUNCtion:IMPedance?
Returned Parameters R, RT
1
, T
2
, LPR, LPRT
3
3.2 FUNCtion:IMPedance:RES:RANGe
Description Resistance measurement range.
Note Values that are not the actual range seing will be evaluated and the range
that contains the value is selected. For example, 110m is greater than the
20m range and less than the 200m range, so 200m is selected.
The return value, and the screen show the range in use and selected.
Command Syntax FUNCtion:IMPedance:RES:RANGe <NR2>
Parameters Range
Values may be sent in many formats: Ex. 20m, 0.020, 2E-2, 0.000002k
Example func:imp:res:range 200E-3
ery Syntax FUNCtion:IMPedance:RES:RANGe?
Returned Parameters <nr3>
1
2841 only
2
see 1
3
see 1
5