User Manual

CHAPTER 12. SYSTEM 31
12.4 SYSTem:LOAD
Description Recall a set of seings from memory
Command Syntax SYSTem:LOAD <nr1>
Parameters The table location number desired
Example syst:load 8
12.5 SYSTem:EXTernalout
Description Output of the comparator results to the handler interface.
Note 2841 only
Command Syntax SYSTem:EXTernalout <str>
Parameters BIN or BCD
The representation of the results.
Example syst:ext bcd
ery Syntax SYSTem:EXTernalout?
Returned Parameters <str> - BIN or BCD
12.6 SYSTem:EOC:MODE
Description End of Conversion indication mode
Command Syntax SYSTem:EOC:MODE <str>
Parameters HOLD or PULSe
Example syst:eoc:mode pulse
ery Syntax SYSTem:EOC:MODE?
Returned Parameters <str>
12.7 SYSTem:EOC:PULSe
Description EOC pulse width
Command Syntax SYSTem:EOC:PULSe <nr2>
Parameters pulse width in seconds
Example SYSTem:EOC:PULS 0.02
ery Syntax SYSTem:EOC:PULSe?
Returned Parameters <nr2>