The following parameters are accessible from the Statistical Display.
3.4.1 OPTIONS
The OPTIONS key allows the user to access the additional statistics options menu.
The measurement options are:
CLEAR: Clear all statistical parameters.
TRIG: Trigger measurement execution.
SAVE DATA: Enabled when a USB flash drive is attached. When activated, resistance and
temperature data is continuously saved to an auto-generated CSV onto the drive. The
key changes to STOP SAVE to terminate the function. SAVE DATA is disabled if no USB
flash drive attached.
Removing the USB flash drive before pressing STOP SAVE will result in the loss of the
recorded data. The CSV is rewritten every time.
3.4.2 EdgeMode
The EdgeMode key allows the user to access two modes.
% (Percent Error): The user can set the nominal value (NOM) and the percent error (%).
If the nominal value is 100 and percentage is 10, the tolerance will be 100±10%. The
instrument will compare the measured value to the tolerance and determine if the DUT
is HI (above tolerance), LO (below tolerance), or IN (within tolerance).
ABS (Absolute): The user can set absolute HIGH and LOW limits. The instrument will
compare the measured values to the absolute limits and determine if the DUT is HI
(above the upper limit), LO (below the lower limit), or IN (within limits).
3.4.3 Status
The Status key allows the user to turn the statistical calculations OFF and ON.
ON: When ON is selected, all functions and buttons are disabled except the trigger and
save screen functions. The instrument makes a statistical measurement at every trigger.
OPTIONS
•CLEAR
•TRIG
•SAVE DATA
EdgeMode
•%
•NOM
•%
•ABS
•Hi
•Li
STATUS
•OFF
•ON