Figure 42 - Boot Screen
If password protection is enabled, the user will have to input the password to operate the unit.
The default password is 2841 See 4.3.4 Password for more details.
2.5 Cable Calibration
The resistance in the 4-terminal cables can be compensated for through the zero adjustment (0
ADJ) function. This should be performed when connecting a new cable to the instrument, when
the environment temperature changes ±5 °C, or when desired.
Complete the following steps to perform the cable calibration:
1. Warm-up the instrument for 30 minutes.
2. Short the 4-terminal connectors together (see Figure 11).
3. Ensure DRIVE+ is aligned with DRIVE- and SENSE+ is aligned with SENSE-.
4. Select a resistance range or AUTO to calibrate for all ranges.
5. Initiate the calibration function. Press 0 ADJ on the front panel.
6. Calibration time is affected by the SPEED, AVERAGE samples, and auto ranging.
Note: If step 3 is not performed, zero adjustment will fail at the 20 mΩ and AUTO range.